Performing Synchronized Cardioversion
This chapter contains information about performing synchronized cardioversion on
a patient with the HP CodeMaster 100.
Refer to “Preparing the Patient” on page 5-5 for information on patient preparation.
7
Performing Cardioversion
Treatment for certain arrhythmias requires synchronizing a defibrillator shock with
the ECG’s R-wave. It is essential that this R-wave is detected to avoid inducing
ventricular fibrillation.
Monitoring During Cardioversion
Using an External Monitor
Whenever possible, Hewlett-Packard recommends that you perform
synchronized cardioversion procedures while directly monitoring the
patient through the multifunction defib electrodes or ECG leads inputs.
If you use an external monitor as the ECG source, ask your biomedical
technician to verify that the CodeMaster 100 defibrillator and external
monitor combination will deliver a synchronized shock within 60 ms of the
peak of the R-wave. (See the
HP M2475B CodeMaster 100 Portable
Defibrillator/Monitor Service Manual
for test procedure.) Use a 1 mV QRS
complex with a QRS width of 40 ms. This performance cannot be
guaranteed with all commercially available monitors.
WA RNI N G
Summary of Contents for Codemaster 100
Page 18: ...Telecom Regulatory Information xviii ...
Page 28: ...xxviii ...
Page 54: ... HWWLQJ FTXDLQWHG Battery Operation 1 26 ...
Page 84: ...6HWXS DQG RQILJXUDWLRQ Configuring the 12 Lead Option 2 30 ...
Page 94: ... HILEULOODWLQJ Defibrillating a Patient 3 10 ...
Page 133: ...8VLQJ WKH HDG 2SWLRQ 12 Lead Keyboard and Display 6 5 Figure 6 2 Characters by Language 6 2 ...
Page 154: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 26 Figure 6 5 A Standard 3x4 ECG 3x4 6 5 ...
Page 178: ...3DFLQJ Using the Pacer 8 8 ...
Page 202: ...7UDQVPLWWLQJ DWD IURP WKH HDG 2SWLRQ Transmitting an Event Summary 10 14 ...
Page 234: ...7URXEOHVKRRWLQJ Service 12 22 ...