Keep hands clear of the paddle electrode edges. Use your thumbs to
depress the Shock buttons on the paddle handles.
10
Grasp the paddle handles, and without removing the paddles from their
holders, press both Shock buttons simultaneously. A brief automatic
recorder run prints the delivered energy test report.
11
Press
Sync
to place the instrument in Sync mode.
12
Verify that the messages
SYNC
and
USE LEADS
appear on the display.
13
Press
Lead Select
once to select Lead I, and verify that the message
USE
LEADS
no longer appears. The message
LEADS OFF
should now appear,
indicating that one or more leads are not connected.
14
Turn the
Energy Select
control to
Off
.
The defibrillator is ready for use if it passes the above checklist.
A more extensive test of defibrillator/monitor functionality can be performed
using the diagnostic service mode described in the
HP M2475B CodeMaster
100 Portable Defibrillator/Monitor Service Manual
.
WA RNI N G
Summary of Contents for Codemaster 100
Page 18: ...Telecom Regulatory Information xviii ...
Page 28: ...xxviii ...
Page 54: ... HWWLQJ FTXDLQWHG Battery Operation 1 26 ...
Page 84: ...6HWXS DQG RQILJXUDWLRQ Configuring the 12 Lead Option 2 30 ...
Page 94: ... HILEULOODWLQJ Defibrillating a Patient 3 10 ...
Page 133: ...8VLQJ WKH HDG 2SWLRQ 12 Lead Keyboard and Display 6 5 Figure 6 2 Characters by Language 6 2 ...
Page 154: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 26 Figure 6 5 A Standard 3x4 ECG 3x4 6 5 ...
Page 178: ...3DFLQJ Using the Pacer 8 8 ...
Page 202: ...7UDQVPLWWLQJ DWD IURP WKH HDG 2SWLRQ Transmitting an Event Summary 10 14 ...
Page 234: ...7URXEOHVKRRWLQJ Service 12 22 ...