7URXEOHVKRRWLQJ
CodeMaster 100
12-3
If the Defibrillator does not Charge
If the defibrillator does not charge, perform the following steps:
1
Verify the proper setting of the
Energy Select
control.
2
If it is correct, follow these steps.
a.
Turn the
Energy Select
control to
Off
, and then back to the desired
energy setting.
b.
Press
Charge
again.
3
If the instrument remains unable to charge, turn the
Energy Select
con-
trol to
Off
and use a backup defibrillator.
4
Press
Review
to print an event summary and keep any ECG strips from
the defibrillator for later evaluation.
5
Alert HP service personnel.
CHECK
BATTERY
7KHEDWWHU\SDFNLVQRWIXOO\LQVHUWHGLQWR
WKHGHILEULOODWRURULWLVIDXOW\
5HLQVHUWWKHEDWWHU\SDFNPDNLQJ
VXUHLWODWFKHV,IWKHPHVVDJH
UHDSSHDUVSHUIRUPDEDWWHU\
FDSDFLW\WHVW
CHECK
SETUP
7KHEDFNXSEDWWHU\IRUWKHVHWXSFRQILJXUD
WLRQPHPRU\LVORZ
&DOOVHUYLFHWRUHSODFHWKHEDFNXS
EDWWHU\
Table 12-2
Defibrillator Messages
(Continued)
Message
Cause
Possible Solutions
Summary of Contents for Codemaster 100
Page 18: ...Telecom Regulatory Information xviii ...
Page 28: ...xxviii ...
Page 54: ... HWWLQJ FTXDLQWHG Battery Operation 1 26 ...
Page 84: ...6HWXS DQG RQILJXUDWLRQ Configuring the 12 Lead Option 2 30 ...
Page 94: ... HILEULOODWLQJ Defibrillating a Patient 3 10 ...
Page 133: ...8VLQJ WKH HDG 2SWLRQ 12 Lead Keyboard and Display 6 5 Figure 6 2 Characters by Language 6 2 ...
Page 154: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 26 Figure 6 5 A Standard 3x4 ECG 3x4 6 5 ...
Page 178: ...3DFLQJ Using the Pacer 8 8 ...
Page 202: ...7UDQVPLWWLQJ DWD IURP WKH HDG 2SWLRQ Transmitting an Event Summary 10 14 ...
Page 234: ...7URXEOHVKRRWLQJ Service 12 22 ...