0DLQWHQDQFH
CodeMaster 100
13-8
7
Turn on the defibrillator and press
Record
to verify the paper is loaded
properly.
Cleaning the Recorder Printhead
If the ECG strip has light or varying density printing, clean the printhead to
remove any possible buildup of paper residue.
To clean the printhead, perform the following steps:
1
Slide the recorder door to the right of the defibrillator. The paper platen
will tilt up.
2
Remove the paper roll.
3
Clean the printhead surface (above the brush) with rubbing alcohol and a
cotton swab.
4
Re-install the paper roll.
5
While holding the recorder door open (to the right), press the platen
down over the paper.
6
Allow the door to close over the platen roller.
7
Press
Record
to test the printer functionality.
Summary of Contents for Codemaster 100
Page 18: ...Telecom Regulatory Information xviii ...
Page 28: ...xxviii ...
Page 54: ... HWWLQJ FTXDLQWHG Battery Operation 1 26 ...
Page 84: ...6HWXS DQG RQILJXUDWLRQ Configuring the 12 Lead Option 2 30 ...
Page 94: ... HILEULOODWLQJ Defibrillating a Patient 3 10 ...
Page 133: ...8VLQJ WKH HDG 2SWLRQ 12 Lead Keyboard and Display 6 5 Figure 6 2 Characters by Language 6 2 ...
Page 154: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 26 Figure 6 5 A Standard 3x4 ECG 3x4 6 5 ...
Page 178: ...3DFLQJ Using the Pacer 8 8 ...
Page 202: ...7UDQVPLWWLQJ DWD IURP WKH HDG 2SWLRQ Transmitting an Event Summary 10 14 ...
Page 234: ...7URXEOHVKRRWLQJ Service 12 22 ...