0DLQWHQDQFH
CodeMaster 100
13-15
Battery Capacity Test
Perform a battery capacity test every 90 days or when the
indicator
(
) flashes after you insert the battery pack into the battery support
system for charging.
1
Insert the battery into charging bay. The
indicator appears
(the
indicator may appear instead).
2
Press the start button(
). The
indicator appears and the
(or
) indicator disappears. Testing the battery takes
approximately 24 hours.
Battery capacity testing is complete when the
PASS
indicator appears.
3
Record the date of the test on the bottom of the battery.
If the
indicator appears, the battery did not pass the test. Remove
the battery from service immediately and notify service personnel.
The battery support system will automatically charge and recondition
batteries during the capacity test.
Dispose of equipment and accessories at the end of their useful life in
accordance with local and national laws.
START
CHARGING
READY
TESTING
CHARGING
READY
WA RNI N G
FAULT
NO TE
NO TE
Summary of Contents for Codemaster 100
Page 18: ...Telecom Regulatory Information xviii ...
Page 28: ...xxviii ...
Page 54: ... HWWLQJ FTXDLQWHG Battery Operation 1 26 ...
Page 84: ...6HWXS DQG RQILJXUDWLRQ Configuring the 12 Lead Option 2 30 ...
Page 94: ... HILEULOODWLQJ Defibrillating a Patient 3 10 ...
Page 133: ...8VLQJ WKH HDG 2SWLRQ 12 Lead Keyboard and Display 6 5 Figure 6 2 Characters by Language 6 2 ...
Page 154: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 26 Figure 6 5 A Standard 3x4 ECG 3x4 6 5 ...
Page 178: ...3DFLQJ Using the Pacer 8 8 ...
Page 202: ...7UDQVPLWWLQJ DWD IURP WKH HDG 2SWLRQ Transmitting an Event Summary 10 14 ...
Page 234: ...7URXEOHVKRRWLQJ Service 12 22 ...