Advisory Mode Default Settings
4-20
4
Advisory Mode Default Settings
The HP CodeMaster 100 contains default settings for use in Advisory mode. Some
of the settings can be altered using the Setup Menu.
The following settings can be changed:
• Default energy select sequence. You can choose to change the sequence from the
factory default settings for a sequence of 200, 200, and 360 Joules, to a sequence
of 200, 300, and 360 Joules. The energy level for the third shock and all subse-
quent shocks will be 360 Joules.
• Post Shock Data: On/Off.
When operating in Advisory mode, the following settings cannot be changed:
• Charge done tone: always sounds.
• CRT alert tones: always sounds.
• Alert tone volume: always at the maximum volume.
• Recorder delay: no delay.
• Recorder Bandwidth: always Monitor bandwidth.
• Automatic Recordings: always off.
• Power on Lead: always PADS/PADDLES.
• Notch Filter: always on.