8VLQJ$X[LOLDU\3RZHU
Using the DC External Power Module
11-6
Testing the External Power Module Installation
To verify that the External Power Module is correctly installed:
1
Connect the external power module to the HP CodeMaster 100 external power
input. Verify that the power indicator on the external power module and the
external power indicator on the HP CodeMaster 100 both appear.
2
Install a battery in the HP CodeMaster 100 and verify that the charging indicator
appears on the External Power Module.
3
Remove the battery and turn the Energy Select/Monitor On control on the HP
CodeMaster 100 to 360J. Press Charge and verify that the defibrillator charged
to 360J in 15 seconds or less.