Defibrillating a Patient
3-6
Defibrillation with Alternate Paddle Sets or Multifunction Defib
Electrodes
The HP CodeMaster 100 will defibrillate with several different paddles/electrode
sets, including:
• Adult/Pediatric, Anterior/Anterior External Paddles
• Adult, Anterior/Posterior External Paddles
• Multifunction Defib Electrodes
Do not switch paddle sets in an environment where water may get in the
paddles/electrodes connector receptacle.
Performing Pediatric Defibrillation
The HP CodeMaster 100 paddle set comes with pediatric paddles. To use the
pediatric paddle set, depress the release latch at the front of the standard external
paddles while pulling forward on the adult paddle surface. This action will remove
the adult paddle contact surface and expose the smaller pediatric contact surface.
Refer to “Defibrillating a Patient” on page 3-1 for defibrillation procedures.
The clinician must select an appropriate energy level for the pediatric
patient. There is no energy limit lockout for the pediatric paddle set.
WA RNI N G
WA RNI N G
Summary of Contents for Codemaster 100
Page 18: ...Telecom Regulatory Information xviii ...
Page 28: ...xxviii ...
Page 54: ... HWWLQJ FTXDLQWHG Battery Operation 1 26 ...
Page 84: ...6HWXS DQG RQILJXUDWLRQ Configuring the 12 Lead Option 2 30 ...
Page 94: ... HILEULOODWLQJ Defibrillating a Patient 3 10 ...
Page 133: ...8VLQJ WKH HDG 2SWLRQ 12 Lead Keyboard and Display 6 5 Figure 6 2 Characters by Language 6 2 ...
Page 154: ...8VLQJ WKH HDG 2SWLRQ Setting Up the Printed Report 6 26 Figure 6 5 A Standard 3x4 ECG 3x4 6 5 ...
Page 178: ...3DFLQJ Using the Pacer 8 8 ...
Page 202: ...7UDQVPLWWLQJ DWD IURP WKH HDG 2SWLRQ Transmitting an Event Summary 10 14 ...
Page 234: ...7URXEOHVKRRWLQJ Service 12 22 ...