External Monitoring
5-14
5
External Monitoring
The ECG output provides an analog 1V/mV ECG signal for connection to a
Hewlett-Packard external monitor. Hewlett-Packard compatible external
monitoring divider cables are as follows.
The ECG output is also compatible with the interface to the HP Viridia
Information Center and other HP central stations.
Do not use the ECG output to synchronize another defibrillator. (The ECG-In
to ECG-Out delay is 35 milliseconds.)
The connection of external equipment may increase leakage currents.
Always request that local safety personnel verify that multiple connected
equipment comply with local regulatory standards before putting such
equipment into service.