Test Sample Chromatograms
Test sample chromatograms
233
Detector Type
TCDw/MUG
Temp 300 DEGREES C
Inlet Type Split only or
split/splitless
Temp
250 DEG C
Operating Mode Split(Purge on)
Purge Time On
0
min
Purge Time Off
0
min
Oven Temp Programmed
Init Temp
100
DEGREES C
Init Time
1
min
Ramp
Rate
10
Fin Temp
150
Fin Time
2
Range
0
COLUMN:
Part No.
19095Z(121)
Dimensions
530
¿
MID; 5 M
Sta Phase
Methyl Silicone
FLOW RATES
Carrier (He)
15 +/• 1
ml/min
Hydrogen
N/A
ml/min
Air
N/A4
ml/min
Makeup (He)
10 +/• 1
ml/min
Split Vent
200 +/• 10
ml/min
Septum Purge
5 +/• 1
ml/min
Reference (He)37 +/• 1
ml/min
SAMPLE:
Type
Capillary
Inj Volume
1
¿
l
Part No. 18740•60900
Composition See Split/FID
4.45
4.78
6.48
ST
.11
.15
.42
.25
3.64
3.85
START
C9
C16
HP 5890 Test Sample Operating
Conditions
TCD-Split Mode Capillary Inlet
Figure 10-9.
Summary of Contents for 5890 Series II
Page 1: ...Reference Manual HP 5890 Series II and HP 5890 Series II Plus ...
Page 8: ...This page intentionally left blank ...
Page 9: ...1 Columns and Fittings ...
Page 31: ...2 Keyboard and Displays ...
Page 42: ...This page intentionally left blank ...
Page 43: ...3 Temperature Control ...
Page 57: ...4 Electronic Flow Sensing ...
Page 66: ...This page intentionally left blank ...
Page 67: ...5 Signal Output ...
Page 98: ...This page intentionally left blank ...
Page 99: ...6 Inlet Systems ...
Page 123: ...7 Detector Systems ...
Page 155: ...8 Preventive Maintenance ...
Page 201: ...9 Chromatographic Troubleshooting ...
Page 222: ...This page intentionally left blank ...
Page 223: ...10 Test Sample Chromatograms ...