Keyboard and Displays
Loading default setpoints
40
Upon pressing
ENTER
, default setpoints are loaded into memory, erasing
setpoints already present. Table 2•1lists resulting HP 5890 default
setpoints.
Function
Default Setpoint
Inj Temp (A & B):
50
^
C, OFF
Det Temp (A & B):
50
^
C, OFF
Oven Temp:
50
^
C, OFF
Oven Max Temp:
400
^
C
Cryo Cooling:
OFF
Cryo Blast
OFF
Equib Time:
3 min
Init Temp:
50
^
C
Init Time:
650 min
Oven Prog Rates:
0
^
C/min
Final Temp:
0
^
C
Final Time:
0 min
Inlet Purge:
ON
Purge Time:
0 min
Detector (A & B):
OFF
Signal 1 Det:
Detector A
Signal 2 Det:
Detector B if BOTH detectors A and
B are installed; otherwise, detector A.
Range (1 & 2):
0
Attn (1 & 2):
0, ON
Zero (1 & 2):
0, ON
HP-IL (INET):
Address- 31
INET:
Global (unconfigured)
Keyboard Lock:
OFF
Timetable
Empty
Inj A Pres
OFF
Inj B Pres
OFF
TCD Sens
HIGH
Oven Track
ON
Constant Flow
OFF
Table 2-1. HP 5890 Default Setpoints
Summary of Contents for 5890 Series II
Page 1: ...Reference Manual HP 5890 Series II and HP 5890 Series II Plus ...
Page 8: ...This page intentionally left blank ...
Page 9: ...1 Columns and Fittings ...
Page 31: ...2 Keyboard and Displays ...
Page 42: ...This page intentionally left blank ...
Page 43: ...3 Temperature Control ...
Page 57: ...4 Electronic Flow Sensing ...
Page 66: ...This page intentionally left blank ...
Page 67: ...5 Signal Output ...
Page 98: ...This page intentionally left blank ...
Page 99: ...6 Inlet Systems ...
Page 123: ...7 Detector Systems ...
Page 155: ...8 Preventive Maintenance ...
Page 201: ...9 Chromatographic Troubleshooting ...
Page 222: ...This page intentionally left blank ...
Page 223: ...10 Test Sample Chromatograms ...