Signal Output
Signal attenuation
74
For analytical information from a detector, proper settings for
RANGE 2
!
( )
and
ATTN 2
!
( )
are determined such that peaks of interest are
on scale
at the integrator or chart recorder: peaks of interest must neither
flat
top
by exceeding the allowed maximum output level, nor be too small to
be measured. Table 5•3lists maximum detector output pro1 volt
at the
+1V
output for each
RANGE 2
!
( )
setpoint value.
FID & NPD (pA)
TCD
(mV, High Gain)
TCD
(mV, Low Gain)
ECD (kHz)
Maximum Detector Signal
Pro1 V Output
0
1.0
}
10
3
25
800
10
1
2.0
}
10
3
50
C
20
2
4.0
}
10
3
C
C
40
3
8.0
}
10
3
C
C
80
4
1.6
}
10
4
C
C
160
5
3.2
}
10
4
C
C
320
6
6.4
}
10
4
C
C
C
7
1.3
}
10
5
C
C
C
8
2.6
}
10
5
C
C
C
9
5.1
}
10
5
C
C
C
10
1.0
}
10
6
C
C
C
11
2.0
}
10
6
C
C
C
12
4.1
}
10
6
C
C
C
13
8.2
}
10
6
C
C
C
RANGE 2
!
( )
Table 5-3. Detector Output Producing a 1 V Signal
Summary of Contents for 5890 Series II
Page 1: ...Reference Manual HP 5890 Series II and HP 5890 Series II Plus ...
Page 8: ...This page intentionally left blank ...
Page 9: ...1 Columns and Fittings ...
Page 31: ...2 Keyboard and Displays ...
Page 42: ...This page intentionally left blank ...
Page 43: ...3 Temperature Control ...
Page 57: ...4 Electronic Flow Sensing ...
Page 66: ...This page intentionally left blank ...
Page 67: ...5 Signal Output ...
Page 98: ...This page intentionally left blank ...
Page 99: ...6 Inlet Systems ...
Page 123: ...7 Detector Systems ...
Page 155: ...8 Preventive Maintenance ...
Page 201: ...9 Chromatographic Troubleshooting ...
Page 222: ...This page intentionally left blank ...
Page 223: ...10 Test Sample Chromatograms ...