Preventive Maintenance
Flame ionization detector (FID)
177
Interconnect
Sprin
g
FID Signal Board Interconnect
Figure 8-13
9. Reassemble the detector cover.
Ignition problems
Before proceeding, make sure that gases are plumbed correctly, the
system is leak•free,flow rates are set correctly, and external lines have
been well purged.
Note:
If Helium is being used as carrier/makeup gas, be aware that
flame lighting problems may occur at very high flow rates ( > 50 ml/min).
If problems arise with He carrier/makeup gas, try using the lowest flow
rate possible, consistent with obtaining good chromatography.
Assuming these things are correct, that the detector is contaminant•free,
and that the ignitor is operational, if ignition problems persist, a
damaged jet may be the cause.
Remove and inspect the jet, looking for galled or scored sealing surfaces,
burrs, a bent or crimped tube, etc. Even if no external damage is visible,
the jet may be damaged internally by overtightening. For this reason, it
Summary of Contents for 5890 Series II
Page 1: ...Reference Manual HP 5890 Series II and HP 5890 Series II Plus ...
Page 8: ...This page intentionally left blank ...
Page 9: ...1 Columns and Fittings ...
Page 31: ...2 Keyboard and Displays ...
Page 42: ...This page intentionally left blank ...
Page 43: ...3 Temperature Control ...
Page 57: ...4 Electronic Flow Sensing ...
Page 66: ...This page intentionally left blank ...
Page 67: ...5 Signal Output ...
Page 98: ...This page intentionally left blank ...
Page 99: ...6 Inlet Systems ...
Page 123: ...7 Detector Systems ...
Page 155: ...8 Preventive Maintenance ...
Page 201: ...9 Chromatographic Troubleshooting ...
Page 222: ...This page intentionally left blank ...
Page 223: ...10 Test Sample Chromatograms ...