Detector Systems
Electron capture detector (ECD)
135
Electron capture detector (ECD)
The effluent gas stream from the detector must be vented to a fume
hood to prevent possible contamination of the laboratory with
radioactive material.
Specific cleaning procedures are provided in Chapter 8, Preventive
Maintenance.
WARNING
Requirements for USA owners
Detector venting must be in conformance with the latest revision of
Title 10, Code Of Federal Regulations, Part 20 (including Appendix B).
This detector is sold under General License: owners may not open the
detector cell or use solvents to clean it. Additional information is
available in the Publication Information For General Licensees, Pub.
No. 43•5953•1798(D).
Owners of this detector must perform a radioactive leak test (wipe test)
at least every six months. The procedure is described in Chapter 8,
Preventive Maintenance.
WARNING
Summary of Contents for 5890 Series II
Page 1: ...Reference Manual HP 5890 Series II and HP 5890 Series II Plus ...
Page 8: ...This page intentionally left blank ...
Page 9: ...1 Columns and Fittings ...
Page 31: ...2 Keyboard and Displays ...
Page 42: ...This page intentionally left blank ...
Page 43: ...3 Temperature Control ...
Page 57: ...4 Electronic Flow Sensing ...
Page 66: ...This page intentionally left blank ...
Page 67: ...5 Signal Output ...
Page 98: ...This page intentionally left blank ...
Page 99: ...6 Inlet Systems ...
Page 123: ...7 Detector Systems ...
Page 155: ...8 Preventive Maintenance ...
Page 201: ...9 Chromatographic Troubleshooting ...
Page 222: ...This page intentionally left blank ...
Page 223: ...10 Test Sample Chromatograms ...