Temperature Control
Valid setpoint ranges
46
Valid setpoint ranges
Table 3•1lists valid setpoint ranges for the 13 keys controlling oven and
heated zone temperatures.
NOTE: TOTAL run time will not exceed 650.00 minutes regardless of values entere
,
, and
.
INIT TIME
RATE
FINAL TIME
-80 to 450
1
^
C
Oven Control
-80 to 450
1
^
C
Oven Control
0 to 650.00
0.01 minute
Oven Control
0 to 70
0.1 /minute
Oven Control
-80 to 450
1
^
C
Oven Control
0 to 650.00
0.01 minute
Oven Control
70 to 450
1
^
C
Oven Control
0 to 200.00
0.01 minute
Oven Control
0 to 400
1
^
C
Zone Control
0 to 400
1
^
C
Zone Control
0 to 400*
1
^
C
Zone Control
0 to 400*
1
^
C
Zone Control
0 to 400
1
^
C
Zone Control
Valid
Setpoint Range
In
Increments Of
Function
Key
OVEN TEMP
INIT TIME
INIT TEMP
FINAL TEMP
INJ A TEMP
INJ B TEMP
FINAL TIME
RATE
DET A TEMP
DET B TEMP
EQUIB TIME
OVEN MAX
AUX TEMP
*The valid setpoint range for a Flame Ionization Detector is 0 to 450
^
C.
Table 3-1. Valid Setpoint Ranges For Temperature Control Keys
Summary of Contents for 5890 Series II
Page 1: ...Reference Manual HP 5890 Series II and HP 5890 Series II Plus ...
Page 8: ...This page intentionally left blank ...
Page 9: ...1 Columns and Fittings ...
Page 31: ...2 Keyboard and Displays ...
Page 42: ...This page intentionally left blank ...
Page 43: ...3 Temperature Control ...
Page 57: ...4 Electronic Flow Sensing ...
Page 66: ...This page intentionally left blank ...
Page 67: ...5 Signal Output ...
Page 98: ...This page intentionally left blank ...
Page 99: ...6 Inlet Systems ...
Page 123: ...7 Detector Systems ...
Page 155: ...8 Preventive Maintenance ...
Page 201: ...9 Chromatographic Troubleshooting ...
Page 222: ...This page intentionally left blank ...
Page 223: ...10 Test Sample Chromatograms ...