Chromatographic Troubleshooting
Peak symptoms
209
Peak symptoms
No peaks
This is usually due to operator error; possibilities include injection on the
wrong column, incorrect signal assignment, attenuation too high (peaks
are present but not visible), a bent syringe needle in an automatic
sampler, etc. Check system parameters for the analysis.
Inverted peaks
This is likely an inappropriate signal assignment definition (e.g.,
B - A
with sample injected on column A) or incorrect polarity with a TCD.
Extra peaks
These are divided into two classes:
Additiona
l peaks appear on the
chart in addition to those expected from the sample.
Ghost
peaks appear
even when no sample is injected (and also appear among the genuine
peaks during a sample run).
1. Peaks appear during a blank run:
These are ghost peaks, usually found during
temperature•programmedruns; the cause is contaminants trapped at
the head of the column at the relatively cool starting temperature.
These are released and chromatographed as column temperature
rises.
Ghost peaks are often observed when a column has been at the
starting temperature for some time. For example, the first few runs in
the morning often contain ghost peaks.
C
Ghost peaks may arise from septum bleed, carrier gas impurities,
and contamination in plumbing by oils, grease, and other
materials. Less commonly, they may be caused by reaction of
Summary of Contents for 5890 Series II
Page 1: ...Reference Manual HP 5890 Series II and HP 5890 Series II Plus ...
Page 8: ...This page intentionally left blank ...
Page 9: ...1 Columns and Fittings ...
Page 31: ...2 Keyboard and Displays ...
Page 42: ...This page intentionally left blank ...
Page 43: ...3 Temperature Control ...
Page 57: ...4 Electronic Flow Sensing ...
Page 66: ...This page intentionally left blank ...
Page 67: ...5 Signal Output ...
Page 98: ...This page intentionally left blank ...
Page 99: ...6 Inlet Systems ...
Page 123: ...7 Detector Systems ...
Page 155: ...8 Preventive Maintenance ...
Page 201: ...9 Chromatographic Troubleshooting ...
Page 222: ...This page intentionally left blank ...
Page 223: ...10 Test Sample Chromatograms ...