Signal Output
File compatibility with data handling devices
97
How to convert HP 339X Integrator workfiles from 5890A to
SERIES II mode:
1. Turn the GC
off
.
2. Follow the previous instructions to set the GC for 5890A mode (use
proper grounding).
3. Download the workfile from the integrator.
4. Turn GC power
off
.
5. Remove the P15 jumper. (Now the GC is in SERIES II mode.)
6. Turn GC power
on
.
7. Add SERIES II setpoints (time table, etc).
8. Store the workfile at the integrator.
Note:
Using
STORE
on the HP 5890 SERIES II, up to 4 integrator
workfiles can be converted with each jumper change sequence.