
Inlet Systems
Split/splitless capillary inlet
115
C
Low Volatility Solute
C
High Volatility Solute
Column
Carrier Gas
Needle
(a)
(b)
(c)
(d)
The Solvent Effect
Figure 6-9
The solvent effect is described in great detail elsewhere: see Grob, K. and
Grob, K., Jr., Journal of Chromatography, 94, page 53 (1974); Grob, K.
and Grob, G., Chromatographia, 5, page 3 (1972).
To reconcentrate sample components via the solvent effect, oven
temperature must be low enough so solvent remains at the head of the
column for a sufficiently long time period. A good guideline is to have the
column 10 to 30
^
C below the solvent boiling point. Table 6•2gives boiling
points for some common solvents, along with suggested initial oven
temperature ranges to use the solvent effect efficiently.
Summary of Contents for 5890 Series II
Page 1: ...Reference Manual HP 5890 Series II and HP 5890 Series II Plus ...
Page 8: ...This page intentionally left blank ...
Page 9: ...1 Columns and Fittings ...
Page 31: ...2 Keyboard and Displays ...
Page 42: ...This page intentionally left blank ...
Page 43: ...3 Temperature Control ...
Page 57: ...4 Electronic Flow Sensing ...
Page 66: ...This page intentionally left blank ...
Page 67: ...5 Signal Output ...
Page 98: ...This page intentionally left blank ...
Page 99: ...6 Inlet Systems ...
Page 123: ...7 Detector Systems ...
Page 155: ...8 Preventive Maintenance ...
Page 201: ...9 Chromatographic Troubleshooting ...
Page 222: ...This page intentionally left blank ...
Page 223: ...10 Test Sample Chromatograms ...