Preventive Maintenance
Flame ionization detector (FID)
172
Jet exchange/replacement
Depending upon the column type to be used, and/or analyses to be
performed, exchanging the jet in an FID may be necessary.
Flame Ionization Detector
Figure 8-9
Note:
The proper jet must be installed prior to column installation.
If switching from packed column operation to capillary operation, the jet
for capillary use
must
be installed prior to column installation.
Depending upon the column type to be installed (packed versus
capillary), the proper jet must be installed. Table 8•1lists available jets: