Columns and Fittings
Inlet inserts
27
Note:
For the liner and insert for an HP Series 530
¿
capillary column, if
the column is already installed, a new insert may not seat properly in the
liner; the column may prevent it from dropping completely into the liner.
If the insert does not drop completely into the liner,
do not
force it
(either the liner or the column may shatter); instead, remove the column,
seat the insert, and then replace the column.
5. Replace the septum and septum retainer nut.
Split/splitless or split-only capillary inlet inserts
Viton O-ring (Preferred)
Viton O-ring
Split Use Splitless Use
Split/Splitless and Split-Only Capillary Inlet and Inserts
Figure 1-10
A specific inlet insert is required, depending upon the particular
sampling mode. Specific sampling modes include:
C
Split, for major•componentanalyses
C
Purged splitless, for trace•componentanalyses
Summary of Contents for 5890 Series II
Page 1: ...Reference Manual HP 5890 Series II and HP 5890 Series II Plus ...
Page 8: ...This page intentionally left blank ...
Page 9: ...1 Columns and Fittings ...
Page 31: ...2 Keyboard and Displays ...
Page 42: ...This page intentionally left blank ...
Page 43: ...3 Temperature Control ...
Page 57: ...4 Electronic Flow Sensing ...
Page 66: ...This page intentionally left blank ...
Page 67: ...5 Signal Output ...
Page 98: ...This page intentionally left blank ...
Page 99: ...6 Inlet Systems ...
Page 123: ...7 Detector Systems ...
Page 155: ...8 Preventive Maintenance ...
Page 201: ...9 Chromatographic Troubleshooting ...
Page 222: ...This page intentionally left blank ...
Page 223: ...10 Test Sample Chromatograms ...