Temperature Control
Fault: messages
52
The message displayed when this occurs is shown in Figure 3•6.
ACTUAL
SETPOINT
WARN:
OVEN
SHUT
OFF
Message, Oven SHUT DOWN
Figure 3-6
The oven remains off until switched on again via the keyboard
(
OVEN TEMP
ON
), unless a
FAULT:
message is displayed (see below,
Fault: messages). Power to the instrument must be switched off, and then
on again to restore operation (setpoints are maintained).
Fault: messages
Figure 3•7shows possible
FAULT:
messages associated with heated zones
or the oven. In general, the following problems are indicated when a
FAULT:
message appears:
C
ADC OFFSET
indicates a problem with one or more electronic
components in circuitry associated with temperature control.
C
LINE SENSE
indicates a problem with AC power to the instrument
(an excessively high source voltage).
C
Any of the
TEMP RDG
messages indicate an inoperative temperature
sensor for the indicated zone or oven.
C
OVEN > MAX + 20
indicates oven temperature exceeds the current
OVEN MAX
setpoint value by more than 20
^
C. Thermal run•awayis
the likely cause.
Note:
In case multiple problems exist simultaneously, press
CLEAR
to
roll
through all message displays.
Summary of Contents for 5890 Series II
Page 1: ...Reference Manual HP 5890 Series II and HP 5890 Series II Plus ...
Page 8: ...This page intentionally left blank ...
Page 9: ...1 Columns and Fittings ...
Page 31: ...2 Keyboard and Displays ...
Page 42: ...This page intentionally left blank ...
Page 43: ...3 Temperature Control ...
Page 57: ...4 Electronic Flow Sensing ...
Page 66: ...This page intentionally left blank ...
Page 67: ...5 Signal Output ...
Page 98: ...This page intentionally left blank ...
Page 99: ...6 Inlet Systems ...
Page 123: ...7 Detector Systems ...
Page 155: ...8 Preventive Maintenance ...
Page 201: ...9 Chromatographic Troubleshooting ...
Page 222: ...This page intentionally left blank ...
Page 223: ...10 Test Sample Chromatograms ...