Automatic
Automatic lets the oscilloscope select the record length for you. The
oscilloscope picks a record length that optimizes the amount of data acquired
and the display update rate. When the sample rate is set to manual and the
record length is set to automatic, the oscilloscope selects the shortest record
length that allows all the data in the waveform record to fill the screen (to a
lower limit of 256 points for a single-wide plug-in and 512 points for a
two-wide plug-in, and to an upper limit of the maximum record length).
When the oscilloscope is in the color graded display, waveform histogram, or
mask testing modes, an optimum number of points for the database is
selected. It is recommended that the auto record length be used for the
above three modes.
Manual
Manual lets you pick a record length from 16 to 16K points with single-wide
plug-ins in the A model mainframes, from 16 to 65K points in the D model
mainframes. (up to 32K or 128K points with a two-wide plug-in, like the
HP 54721A; or up to 64K or 256K points with a four-wide plug-in, like the
HP 54722A). Use the knob, arrow keys, or keypad to select a record length.
Remember that sample rate and record length work together. If you combine
a small record length with a high sample rate, you will have a very fast
throughput, but very little data in the record.
Acquisition Menu
Record Length
3–15
Summary of Contents for 54710A
Page 7: ...vi ...
Page 9: ...viii ...
Page 16: ...Glossary Index Contents Contents 7 ...
Page 17: ...Contents 8 ...
Page 65: ...2 16 ...
Page 86: ...4 Applications ...
Page 96: ...6 Display 6 4 Scale 6 5 Offset 6 6 Input 6 6 Probe 6 7 Calibrate 6 10 Channel Menu ...
Page 98: ...Channel menu and menu map for the HP 54712A plug in Figure 6 1 Channel Menu 6 3 ...
Page 110: ...7 Thresholds 7 4 Top Base 7 6 Define time 7 8 Statistics 7 9 Define Measure Menu ...
Page 112: ...Define measure menu and menu map Figure 7 2 Define Measure Menu 7 3 ...
Page 143: ...8 22 ...
Page 163: ...9 20 ...
Page 164: ...10 Messages ...
Page 178: ...12 Function 12 3 Define Function 12 4 Display 12 7 Math Menu ...
Page 186: ...13 Measurements ...
Page 245: ...13 60 ...
Page 246: ...14 Setup Memory 14 3 Save 14 3 Recall 14 4 Default Setups 14 4 Setup Menu ...
Page 261: ...15 10 ...
Page 274: ...17 Scale 17 3 Position 17 3 Reference 17 4 Windowing 17 5 Time Base Menu ...
Page 282: ...Trigger menu and menu map Figure 18 1 Trigger Menu 18 3 ...
Page 319: ...19 18 ...
Page 322: ...20 Waveform 20 3 Pixel 20 6 Waveform Menu ...
Page 346: ...Limit test menu and menu map Figure 22 2 Limit Test Menu 22 3 ...
Page 357: ...22 14 ...
Page 360: ...Mask menu map Figure 23 1 Mask Menu 23 3 ...
Page 370: ...Figure 23 3 Figure 23 4 Mask Menu Edit Mask 23 13 ...
Page 407: ...Index 10 ...
Page 409: ......