Sampling Rate
Sampling rate and waveform record length (also called waveform memory)
are very closely tied together. Sampling rate determines how often the
oscilloscope samples the signal you are measuring, and waveform record
length is the amount of memory the oscilloscope must fill up before it
updates the measurement results.
Usually, you use a high sample rate, so there are plenty of data points to
better reconstruct the original waveform. For fastest throughput, you use a
short waveform record because the oscilloscope must fill up the waveform
record before it updates the measurement results. The higher the sample
rate, the faster any waveform record fills up. A long waveform record slows
down the throughput of the oscilloscope. As a result, you must often make a
trade-off between sample rate and record length, that also results in
measurement speed and accurate reconstruction of the waveform.
If you set the sample rate and record length to automatic, the oscilloscope
picks an optimum sample rate and record length that gives the best waveform
reconstruction and measurement speed for the time base scale setting you
selected. However, you may want to make the trade-off decisions yourself by
selecting a sample rate and record length that fits your application. You may
prefer to sacrifice measurement throughput by selecting a high sample rate
and a long record length, or you may want to have a high sample rate and a
short record length to maximize the measurement throughput.
There are some points you should remember when you are mixing plug-ins,
(using both single-wide plug-ins and a two-wide plug-in). The maximum
sampling rate is 2 GSa/s per plug-in slot. A two-wide plug-in, like the
HP 54721A, uses two plug-in slots which allows a maximum sampling rate of
4 GSa/s. A four-wide plug-in, like the HP 54722A, uses four plug-in slots
which allows a maximum sampling rate of 8 GSa/s. The selected sample rate
is displayed in the acquisition menu and in the status area of the screen.
(The upper left corner of the screen is the status area.)
Acquisition Menu
Sampling Rate
3–12
Summary of Contents for 54710A
Page 7: ...vi ...
Page 9: ...viii ...
Page 16: ...Glossary Index Contents Contents 7 ...
Page 17: ...Contents 8 ...
Page 65: ...2 16 ...
Page 86: ...4 Applications ...
Page 96: ...6 Display 6 4 Scale 6 5 Offset 6 6 Input 6 6 Probe 6 7 Calibrate 6 10 Channel Menu ...
Page 98: ...Channel menu and menu map for the HP 54712A plug in Figure 6 1 Channel Menu 6 3 ...
Page 110: ...7 Thresholds 7 4 Top Base 7 6 Define time 7 8 Statistics 7 9 Define Measure Menu ...
Page 112: ...Define measure menu and menu map Figure 7 2 Define Measure Menu 7 3 ...
Page 143: ...8 22 ...
Page 163: ...9 20 ...
Page 164: ...10 Messages ...
Page 178: ...12 Function 12 3 Define Function 12 4 Display 12 7 Math Menu ...
Page 186: ...13 Measurements ...
Page 245: ...13 60 ...
Page 246: ...14 Setup Memory 14 3 Save 14 3 Recall 14 4 Default Setups 14 4 Setup Menu ...
Page 261: ...15 10 ...
Page 274: ...17 Scale 17 3 Position 17 3 Reference 17 4 Windowing 17 5 Time Base Menu ...
Page 282: ...Trigger menu and menu map Figure 18 1 Trigger Menu 18 3 ...
Page 319: ...19 18 ...
Page 322: ...20 Waveform 20 3 Pixel 20 6 Waveform Menu ...
Page 346: ...Limit test menu and menu map Figure 22 2 Limit Test Menu 22 3 ...
Page 357: ...22 14 ...
Page 360: ...Mask menu map Figure 23 1 Mask Menu 23 3 ...
Page 370: ...Figure 23 3 Figure 23 4 Mask Menu Edit Mask 23 13 ...
Page 407: ...Index 10 ...
Page 409: ......