![HP 54710A User'S Reference Manual Download Page 232](http://html.mh-extra.com/html/hp/54710a/54710a_users-reference-manual_166178232.webp)
Advantages of real-time sampling include the following:
•
Usable for single-shot measurements
•
Acquires a complete record of each edge on each trigger.
Disadvantages of real-time sampling include the following:
•
Sampling rate limits the bandwidth
•
Memory depth limits the time resolution
•
Slow sweep speeds increase the possibility of aliasing (see "Bandwidth,
Sampling Rate, and Reconstruction" in this section).
Equivalent-Time Sampling
In equivalent-time sampling, the signal is constantly sampled and digitized at
a rate determined by the oscilloscope’s sampling clock (see figure 21).
Whenever a trigger event occurs, the oscilloscope measures the time from
the trigger event to the next sample. Knowing this time relationship allows
the oscilloscope to assign the samples to the correct time relative to the
trigger. After enough acquisition cycles have occurred, the waveform is
completely defined. Equivalent-time sampling requires a repetitive signal.
Because the sampling phase is random with respect to the input signal,
equivalent-time sampling is very insensitive to aliasing on repetitive signals.
Measurements
Time-interval measurements
13–47
Summary of Contents for 54710A
Page 7: ...vi ...
Page 9: ...viii ...
Page 16: ...Glossary Index Contents Contents 7 ...
Page 17: ...Contents 8 ...
Page 65: ...2 16 ...
Page 86: ...4 Applications ...
Page 96: ...6 Display 6 4 Scale 6 5 Offset 6 6 Input 6 6 Probe 6 7 Calibrate 6 10 Channel Menu ...
Page 98: ...Channel menu and menu map for the HP 54712A plug in Figure 6 1 Channel Menu 6 3 ...
Page 110: ...7 Thresholds 7 4 Top Base 7 6 Define time 7 8 Statistics 7 9 Define Measure Menu ...
Page 112: ...Define measure menu and menu map Figure 7 2 Define Measure Menu 7 3 ...
Page 143: ...8 22 ...
Page 163: ...9 20 ...
Page 164: ...10 Messages ...
Page 178: ...12 Function 12 3 Define Function 12 4 Display 12 7 Math Menu ...
Page 186: ...13 Measurements ...
Page 245: ...13 60 ...
Page 246: ...14 Setup Memory 14 3 Save 14 3 Recall 14 4 Default Setups 14 4 Setup Menu ...
Page 261: ...15 10 ...
Page 274: ...17 Scale 17 3 Position 17 3 Reference 17 4 Windowing 17 5 Time Base Menu ...
Page 282: ...Trigger menu and menu map Figure 18 1 Trigger Menu 18 3 ...
Page 319: ...19 18 ...
Page 322: ...20 Waveform 20 3 Pixel 20 6 Waveform Menu ...
Page 346: ...Limit test menu and menu map Figure 22 2 Limit Test Menu 22 3 ...
Page 357: ...22 14 ...
Page 360: ...Mask menu map Figure 23 1 Mask Menu 23 3 ...
Page 370: ...Figure 23 3 Figure 23 4 Mask Menu Edit Mask 23 13 ...
Page 407: ...Index 10 ...
Page 409: ......