The maximum sample rate allows you to have the maximum single-shot
bandwidth.
The real-time sampling mode is intended to capture events that happen
either once or infrequently. To accurately display infrequently occurring
events, you need to capture and display as much data as possible with a
single trigger. For example, use the real-time sampling mode for laser
applications, high-speed glitch capture in digital systems, eye diagrams, or
when looking for transient events.
Equivalent time
In the equivalent-time sampling mode, the waveform is reconstructed from
data points that are acquired from multiple trigger events. The
equivalent-time sampling mode has the following features:
•
8 bits of vertical resolution
•
Averaging of acquired data points to achieve vertical resolutions of up to
12 bits.
•
Maximum sampling rate of 500 MSa/s on each plug-in slot
•
Maximum system bandwidth
•
Ability to turn on or off averaging
•
Selectable record length
•
Maximum of 32K points for any equivalent-time waveform.
The equivalent-time sampling mode is intended for use on repetitive signals.
For repetitive signals, you can use the data acquired from multiple triggers to
reconstruct the waveform, which gives you a higher system bandwidth.
Acquisition Menu
Sampling Mode
3–5
Summary of Contents for 54710A
Page 7: ...vi ...
Page 9: ...viii ...
Page 16: ...Glossary Index Contents Contents 7 ...
Page 17: ...Contents 8 ...
Page 65: ...2 16 ...
Page 86: ...4 Applications ...
Page 96: ...6 Display 6 4 Scale 6 5 Offset 6 6 Input 6 6 Probe 6 7 Calibrate 6 10 Channel Menu ...
Page 98: ...Channel menu and menu map for the HP 54712A plug in Figure 6 1 Channel Menu 6 3 ...
Page 110: ...7 Thresholds 7 4 Top Base 7 6 Define time 7 8 Statistics 7 9 Define Measure Menu ...
Page 112: ...Define measure menu and menu map Figure 7 2 Define Measure Menu 7 3 ...
Page 143: ...8 22 ...
Page 163: ...9 20 ...
Page 164: ...10 Messages ...
Page 178: ...12 Function 12 3 Define Function 12 4 Display 12 7 Math Menu ...
Page 186: ...13 Measurements ...
Page 245: ...13 60 ...
Page 246: ...14 Setup Memory 14 3 Save 14 3 Recall 14 4 Default Setups 14 4 Setup Menu ...
Page 261: ...15 10 ...
Page 274: ...17 Scale 17 3 Position 17 3 Reference 17 4 Windowing 17 5 Time Base Menu ...
Page 282: ...Trigger menu and menu map Figure 18 1 Trigger Menu 18 3 ...
Page 319: ...19 18 ...
Page 322: ...20 Waveform 20 3 Pixel 20 6 Waveform Menu ...
Page 346: ...Limit test menu and menu map Figure 22 2 Limit Test Menu 22 3 ...
Page 357: ...22 14 ...
Page 360: ...Mask menu map Figure 23 1 Mask Menu 23 3 ...
Page 370: ...Figure 23 3 Figure 23 4 Mask Menu Edit Mask 23 13 ...
Page 407: ...Index 10 ...
Page 409: ......