![HP 54710A User'S Reference Manual Download Page 169](http://html.mh-extra.com/html/hp/54710a/54710a_users-reference-manual_166178169.webp)
Probe attenuation (or gain) exceeds calibration limits
The automatic probe cal cannot generate the cal factors to compensate for
the gain or attenuation of the probe. The attenuation factor limits are
−
80 dB
to
−
120 dB..
Setup defaulted
This normally occurs when you ask for setup to be defaulted. At power on,
the setups are defaulted if a memory error was detected or had occurred
prior to power down. The setup is also defaulted when new system firmware
is installed or when the memory is declassified in the utility menu.
System firmware error occurred
This is a defect diagnostic that should be reported to Hewlett-Packard
service. Normally, the address where the error occurred is also displayed.
Record the address to help in servicing the instrument. The instrument is
still operable.
System load failed: System firmware is unchanged
This can be caused by either a corrupt disk file or a hardware failure. Try
another system disk file to narrow the problem to the file or the hardware. If
it is a hardware failure, call Hewlett-Packard for service.
Messages
10-6
Summary of Contents for 54710A
Page 7: ...vi ...
Page 9: ...viii ...
Page 16: ...Glossary Index Contents Contents 7 ...
Page 17: ...Contents 8 ...
Page 65: ...2 16 ...
Page 86: ...4 Applications ...
Page 96: ...6 Display 6 4 Scale 6 5 Offset 6 6 Input 6 6 Probe 6 7 Calibrate 6 10 Channel Menu ...
Page 98: ...Channel menu and menu map for the HP 54712A plug in Figure 6 1 Channel Menu 6 3 ...
Page 110: ...7 Thresholds 7 4 Top Base 7 6 Define time 7 8 Statistics 7 9 Define Measure Menu ...
Page 112: ...Define measure menu and menu map Figure 7 2 Define Measure Menu 7 3 ...
Page 143: ...8 22 ...
Page 163: ...9 20 ...
Page 164: ...10 Messages ...
Page 178: ...12 Function 12 3 Define Function 12 4 Display 12 7 Math Menu ...
Page 186: ...13 Measurements ...
Page 245: ...13 60 ...
Page 246: ...14 Setup Memory 14 3 Save 14 3 Recall 14 4 Default Setups 14 4 Setup Menu ...
Page 261: ...15 10 ...
Page 274: ...17 Scale 17 3 Position 17 3 Reference 17 4 Windowing 17 5 Time Base Menu ...
Page 282: ...Trigger menu and menu map Figure 18 1 Trigger Menu 18 3 ...
Page 319: ...19 18 ...
Page 322: ...20 Waveform 20 3 Pixel 20 6 Waveform Menu ...
Page 346: ...Limit test menu and menu map Figure 22 2 Limit Test Menu 22 3 ...
Page 357: ...22 14 ...
Page 360: ...Mask menu map Figure 23 1 Mask Menu 23 3 ...
Page 370: ...Figure 23 3 Figure 23 4 Mask Menu Edit Mask 23 13 ...
Page 407: ...Index 10 ...
Page 409: ......