Calibrate to best accuracy
Calibrate to best accuracy allows you to achieve the one percent accuracy
specifications. A best accuracy calibration is valid for a specific plug-in, in a
specific slot, in a specific mainframe. It is valid as long as the current
temperature
∆
and plug-in best accuracy temperature
∆
are within
±
5
degrees, and no more than 24 hours has elapsed since the best accuracy
calibration was performed.
The calibration factors for a best accuracy calibration are stored in the
mainframe’s memory. The calibration factors are retained until another best
accuracy calibration is performed in this slot.
If you perform another best accuracy calibration on that plug-in in a different
slot or mainframe, when you bring the plug-in back into the original slot, and
as long as another best accuracy calibration was not performed in that slot,
the original best accuracy calibration factors are retained.
The best accuracy calibration factors for a two-wide plug-in are stored in the
left slot. For example, if you performed a best accuracy calibration on a
two-wide plug-in in slots 2 and 3, the best accuracy calibration factors are
stored in the memory behind slot two. You can remove the two-wide plug-in
and install a single-wide plug-in in slot 3. You can perform a best accuracy
calibration on the single-wide plug-in in slot 3 without affecting the best
accuracy calibration factors for the two-wide plug-in that are stored behind
slot 2.
Also, there is a best cal protect switch on the rear panel of the oscilloscope.
The normal setting for this switch is unprotected so that you can perform a
best accuracy calibration at any time without having to access the rear panel
of the oscilloscope.
To perform a best accuracy calibration
Press the Calibrate to best accuracy softkey. Then, follow the instructions on
the display.
Channel Menu
Calibrate
6–11
Summary of Contents for 54710A
Page 7: ...vi ...
Page 9: ...viii ...
Page 16: ...Glossary Index Contents Contents 7 ...
Page 17: ...Contents 8 ...
Page 65: ...2 16 ...
Page 86: ...4 Applications ...
Page 96: ...6 Display 6 4 Scale 6 5 Offset 6 6 Input 6 6 Probe 6 7 Calibrate 6 10 Channel Menu ...
Page 98: ...Channel menu and menu map for the HP 54712A plug in Figure 6 1 Channel Menu 6 3 ...
Page 110: ...7 Thresholds 7 4 Top Base 7 6 Define time 7 8 Statistics 7 9 Define Measure Menu ...
Page 112: ...Define measure menu and menu map Figure 7 2 Define Measure Menu 7 3 ...
Page 143: ...8 22 ...
Page 163: ...9 20 ...
Page 164: ...10 Messages ...
Page 178: ...12 Function 12 3 Define Function 12 4 Display 12 7 Math Menu ...
Page 186: ...13 Measurements ...
Page 245: ...13 60 ...
Page 246: ...14 Setup Memory 14 3 Save 14 3 Recall 14 4 Default Setups 14 4 Setup Menu ...
Page 261: ...15 10 ...
Page 274: ...17 Scale 17 3 Position 17 3 Reference 17 4 Windowing 17 5 Time Base Menu ...
Page 282: ...Trigger menu and menu map Figure 18 1 Trigger Menu 18 3 ...
Page 319: ...19 18 ...
Page 322: ...20 Waveform 20 3 Pixel 20 6 Waveform Menu ...
Page 346: ...Limit test menu and menu map Figure 22 2 Limit Test Menu 22 3 ...
Page 357: ...22 14 ...
Page 360: ...Mask menu map Figure 23 1 Mask Menu 23 3 ...
Page 370: ...Figure 23 3 Figure 23 4 Mask Menu Edit Mask 23 13 ...
Page 407: ...Index 10 ...
Page 409: ......