Example
Other devices
The mainframe’s CAL signal is a voltage source. You cannot use it to
calibrate to the probe tip when units are set to Ampere, Watt, or unknown.
Instead, you set the external gain and external offset to compensate for the
actual characteristics of your probe or device. If you do not know its actual
characteristics, you can refer to the specifications that came with the probe
or device. The following steps show you how to do this.
1
Set
Atten units
to
ratio
.
2
Set
Probe atten
to
1:1
.
3
Set
Units
to
Ampere
,
Watt
, or
unknown
.
4
Set
Ext gain
to the actual gain characteristics of your probe or device.
5
Set
Ext offset
to the offset that is introduced by your probe or device.
Calibrate probe
Connect a voltage probe from the plug-in to the CAL signal on the mainframe,
then press the Calibrate probe softkey. The oscilloscope calibrates to the tip
of the probe by setting the probe attenuation to the actual attenuation ratio
of the probe. The oscilloscope also automatically compensates for any offset
that the probe may introduce.
Channel Menu
Probe
6–9
Summary of Contents for 54710A
Page 7: ...vi ...
Page 9: ...viii ...
Page 16: ...Glossary Index Contents Contents 7 ...
Page 17: ...Contents 8 ...
Page 65: ...2 16 ...
Page 86: ...4 Applications ...
Page 96: ...6 Display 6 4 Scale 6 5 Offset 6 6 Input 6 6 Probe 6 7 Calibrate 6 10 Channel Menu ...
Page 98: ...Channel menu and menu map for the HP 54712A plug in Figure 6 1 Channel Menu 6 3 ...
Page 110: ...7 Thresholds 7 4 Top Base 7 6 Define time 7 8 Statistics 7 9 Define Measure Menu ...
Page 112: ...Define measure menu and menu map Figure 7 2 Define Measure Menu 7 3 ...
Page 143: ...8 22 ...
Page 163: ...9 20 ...
Page 164: ...10 Messages ...
Page 178: ...12 Function 12 3 Define Function 12 4 Display 12 7 Math Menu ...
Page 186: ...13 Measurements ...
Page 245: ...13 60 ...
Page 246: ...14 Setup Memory 14 3 Save 14 3 Recall 14 4 Default Setups 14 4 Setup Menu ...
Page 261: ...15 10 ...
Page 274: ...17 Scale 17 3 Position 17 3 Reference 17 4 Windowing 17 5 Time Base Menu ...
Page 282: ...Trigger menu and menu map Figure 18 1 Trigger Menu 18 3 ...
Page 319: ...19 18 ...
Page 322: ...20 Waveform 20 3 Pixel 20 6 Waveform Menu ...
Page 346: ...Limit test menu and menu map Figure 22 2 Limit Test Menu 22 3 ...
Page 357: ...22 14 ...
Page 360: ...Mask menu map Figure 23 1 Mask Menu 23 3 ...
Page 370: ...Figure 23 3 Figure 23 4 Mask Menu Edit Mask 23 13 ...
Page 407: ...Index 10 ...
Page 409: ......