Store screen
Store screen determines what the oscilloscope does to the data on the screen
on a failure. The screen data is a pixel dump of the entire screen area. The
choices are off, pixel memory, printer, disk. The screen image allows you to
see what the display looked like at the time of the failure , but you cannot
rescale it later like you can with waveform memories or functions. If the help
menu is displayed on the screen at the time of the failure, then the help menu
is the screen image that is stored.
Off
Off means that the an image copy of the entire screen area is not
stored, printed, or retained.
Pixel memory
Pixel memory stores an image copy of the entire screen
area to the pixel memory; the same pixel memory that is accessed by the
Waveform menu. Additional failures are added to the composite data
stored in the pixel memory. The only way to clear the pixel memory is
from the Waveform menu.
The value of the pixel memory selection is that you can save the waveform
area from multiple failures to the pixel memory and have a composite view of
the failures.
See also
Chapter 20, "Waveform menu," for additional information on the pixel
memory.
Limit Test Menu
Fail Action
22–12
Summary of Contents for 54710A
Page 7: ...vi ...
Page 9: ...viii ...
Page 16: ...Glossary Index Contents Contents 7 ...
Page 17: ...Contents 8 ...
Page 65: ...2 16 ...
Page 86: ...4 Applications ...
Page 96: ...6 Display 6 4 Scale 6 5 Offset 6 6 Input 6 6 Probe 6 7 Calibrate 6 10 Channel Menu ...
Page 98: ...Channel menu and menu map for the HP 54712A plug in Figure 6 1 Channel Menu 6 3 ...
Page 110: ...7 Thresholds 7 4 Top Base 7 6 Define time 7 8 Statistics 7 9 Define Measure Menu ...
Page 112: ...Define measure menu and menu map Figure 7 2 Define Measure Menu 7 3 ...
Page 143: ...8 22 ...
Page 163: ...9 20 ...
Page 164: ...10 Messages ...
Page 178: ...12 Function 12 3 Define Function 12 4 Display 12 7 Math Menu ...
Page 186: ...13 Measurements ...
Page 245: ...13 60 ...
Page 246: ...14 Setup Memory 14 3 Save 14 3 Recall 14 4 Default Setups 14 4 Setup Menu ...
Page 261: ...15 10 ...
Page 274: ...17 Scale 17 3 Position 17 3 Reference 17 4 Windowing 17 5 Time Base Menu ...
Page 282: ...Trigger menu and menu map Figure 18 1 Trigger Menu 18 3 ...
Page 319: ...19 18 ...
Page 322: ...20 Waveform 20 3 Pixel 20 6 Waveform Menu ...
Page 346: ...Limit test menu and menu map Figure 22 2 Limit Test Menu 22 3 ...
Page 357: ...22 14 ...
Page 360: ...Mask menu map Figure 23 1 Mask Menu 23 3 ...
Page 370: ...Figure 23 3 Figure 23 4 Mask Menu Edit Mask 23 13 ...
Page 407: ...Index 10 ...
Page 409: ......