Normal indicates that the plug-in is not calibrated for best accuracy. Either
the best accuracy calibration factors were cleared from the memory in the
mainframe, or that a best accuracy calibration has not yet been performed on
this plug-in in this slot. But calibrated does indicate that the normal plug-in
calibration is still valid for this plug-in which gives you typically a 3 percent
vertical accuracy characteristic. All you have to do to gain back the 1 percent
best accuracy specifications is perform a best accuracy calibration again.
To perform a best accuracy calibration
1
Press the Channel key on the plug-in.
2
Press the Calibrate...softkey.
3
Press the Calibrate to best accuracy softkey.
4
Follow the instructions on the display.
Uncalibrated indicates that the calibration factors in the plug-in are set to the
default state. See the Service Guide that is supplied with the plug-in for
calibration information.
Best accuracy
∆
Temp
Indicates the temperature change from the
temperature that the last best accuracy calibration was performed at.
Output
Output lets you select the output signal for the front-panel CAL signal on the
mainframe. The output choices are dc, 2-kHz square wave and 500 kHz
square wave. The CAL signal is used for mainframe calibration, plug-in
calibration, probe calibration, best accuracy calibration, skew calibration, and
as a demonstration signal.
•
The dc output is setable from
−
2.5 Vdc to
+
2.5 Vdc into 50
Ω
.
•
The 2-kHz and 500-kHz square waves are 1 Vp-p into 50
Ω
with about
1.6-ns edges.
Utility Menu
Calibrate
19–13
Summary of Contents for 54710A
Page 7: ...vi ...
Page 9: ...viii ...
Page 16: ...Glossary Index Contents Contents 7 ...
Page 17: ...Contents 8 ...
Page 65: ...2 16 ...
Page 86: ...4 Applications ...
Page 96: ...6 Display 6 4 Scale 6 5 Offset 6 6 Input 6 6 Probe 6 7 Calibrate 6 10 Channel Menu ...
Page 98: ...Channel menu and menu map for the HP 54712A plug in Figure 6 1 Channel Menu 6 3 ...
Page 110: ...7 Thresholds 7 4 Top Base 7 6 Define time 7 8 Statistics 7 9 Define Measure Menu ...
Page 112: ...Define measure menu and menu map Figure 7 2 Define Measure Menu 7 3 ...
Page 143: ...8 22 ...
Page 163: ...9 20 ...
Page 164: ...10 Messages ...
Page 178: ...12 Function 12 3 Define Function 12 4 Display 12 7 Math Menu ...
Page 186: ...13 Measurements ...
Page 245: ...13 60 ...
Page 246: ...14 Setup Memory 14 3 Save 14 3 Recall 14 4 Default Setups 14 4 Setup Menu ...
Page 261: ...15 10 ...
Page 274: ...17 Scale 17 3 Position 17 3 Reference 17 4 Windowing 17 5 Time Base Menu ...
Page 282: ...Trigger menu and menu map Figure 18 1 Trigger Menu 18 3 ...
Page 319: ...19 18 ...
Page 322: ...20 Waveform 20 3 Pixel 20 6 Waveform Menu ...
Page 346: ...Limit test menu and menu map Figure 22 2 Limit Test Menu 22 3 ...
Page 357: ...22 14 ...
Page 360: ...Mask menu map Figure 23 1 Mask Menu 23 3 ...
Page 370: ...Figure 23 3 Figure 23 4 Mask Menu Edit Mask 23 13 ...
Page 407: ...Index 10 ...
Page 409: ......