![HP 54710A User'S Reference Manual Download Page 267](http://html.mh-extra.com/html/hp/54710a/54710a_users-reference-manual_166178267.webp)
Fast Fourier
Transforms
(FFT)
Frequency Specifications
(HP 54710A/20A in Real Time Mode)
Plug-in
Bandwidth
(-3 dB)
Maximum
Sample Rate
Record
Length
Maximum
Freq Span
Max Resolution
at Max Frequency
54711A
>1.5 GHz
1
2 GSa/s
16-16,384
2
1 GHz
122 kHz
3
54712A
>1.1 GHz
2 GSa/s
16-16,384
2
1 GHz
122 kHz
3
54713A
>500 MHz
2 GSa/s
16-16,384
2
1 GHz
122 kHz
3
54714A
>500 MHz
2 GSa/s
16-8,192
2
1 GHz
244 kHz
3
54721A
>1.1 GHz
4 GSa/s
16-32,768
2 GHz
122 kHz
54722A
>1.5 GHz
1
8 GSa/s
16-32,768
4 GHz
244 kHz
1. 2.0 GHz in the HP 54710D and 54720D mainframes
2. 16-32,768 in the HP 54710D and 54720D mainframes
3. 61 kHz in the HP 54710D and 54720D mainframes
Span
sample rate / 2
Resolution
sample rate / record length
Frequency Accuracy
±
[(0.5
×
resolution)
+
(0.00007
×
signal
frequency)]
Magnitude Specifications
Magnitude Accuracy (Flattop window)
0.26 dB (3%) Near dc,
−
3
dB at maximum bandwidth
Signal-to-noise Ratio
55 to 65 dB (typical).
Controls
Span
Sets the maximum frequency
Resolution
Sets the spacing between points in the frequency domain
Magnify Controls
For zooming in on a portion of the frequency
record up to an expansion factor of 200. The magnify controls are:
magnify on/off, magnify span, and center frequency.
Vertical Scaling
The vertical scale is adjustable in dBm, a
logarithmic scale. You can adjust sensitivity and offset.
Window
There are three windows for reducing leakage and
enhancing certain characteristics in the frequency domain.
Rectangular
(window
off)
Use for transients and signals where
there are an integral number of cycles in the time record.
Hanning
Use for frequency accuracy and general purpose use.
Specifications and Characteristics
Characteristics
16–6
Summary of Contents for 54710A
Page 7: ...vi ...
Page 9: ...viii ...
Page 16: ...Glossary Index Contents Contents 7 ...
Page 17: ...Contents 8 ...
Page 65: ...2 16 ...
Page 86: ...4 Applications ...
Page 96: ...6 Display 6 4 Scale 6 5 Offset 6 6 Input 6 6 Probe 6 7 Calibrate 6 10 Channel Menu ...
Page 98: ...Channel menu and menu map for the HP 54712A plug in Figure 6 1 Channel Menu 6 3 ...
Page 110: ...7 Thresholds 7 4 Top Base 7 6 Define time 7 8 Statistics 7 9 Define Measure Menu ...
Page 112: ...Define measure menu and menu map Figure 7 2 Define Measure Menu 7 3 ...
Page 143: ...8 22 ...
Page 163: ...9 20 ...
Page 164: ...10 Messages ...
Page 178: ...12 Function 12 3 Define Function 12 4 Display 12 7 Math Menu ...
Page 186: ...13 Measurements ...
Page 245: ...13 60 ...
Page 246: ...14 Setup Memory 14 3 Save 14 3 Recall 14 4 Default Setups 14 4 Setup Menu ...
Page 261: ...15 10 ...
Page 274: ...17 Scale 17 3 Position 17 3 Reference 17 4 Windowing 17 5 Time Base Menu ...
Page 282: ...Trigger menu and menu map Figure 18 1 Trigger Menu 18 3 ...
Page 319: ...19 18 ...
Page 322: ...20 Waveform 20 3 Pixel 20 6 Waveform Menu ...
Page 346: ...Limit test menu and menu map Figure 22 2 Limit Test Menu 22 3 ...
Page 357: ...22 14 ...
Page 360: ...Mask menu map Figure 23 1 Mask Menu 23 3 ...
Page 370: ...Figure 23 3 Figure 23 4 Mask Menu Edit Mask 23 13 ...
Page 407: ...Index 10 ...
Page 409: ......