Summary of dc Errors
You can minimize the effects of dc errors on time-interval measurements by
observing these guidelines when feasible:
•
Measure between identical edges—same slew rate, amplitude, and offset.
•
Use the same dc level to reference each endpoint of the interval.
•
Reference the measurement to the fastest-slewing portion of each edge.
•
Operate the oscilloscope in a mode that gives the best vertical resolution.
Signal edges having faster slew rates minimize the effects of dc errors on the
time interval measurement.
Dynamic Response Errors
Whenever the input voltage of an oscilloscope changes with respect to time,
there is some transient error in the oscilloscope’s reproduction of the input
signal.
The probe also has some non ideal transient response that contributes to the
overall system error. Additionally, the probe and the circuit under test form
a circuit that behaves differently from the circuit without the probe; the
signal at the probe tip is not the same as the signal that is present when the
probe is removed. Probe loading often has a more significant effect on the
measurement than the transient response of the oscilloscope, and must be
included in the analysis of time-interval measurement errors. See "Probe
Capacitive Loading" for more information.
At the very high bandwidth of the HP 54720 oscilloscope, the transient
response of 50
Ω
coaxial cables can be significant. You must use good quality
cables and keep them as short as possible for best accuracy.
Measurements
Time-interval measurements
13–37
Summary of Contents for 54710A
Page 7: ...vi ...
Page 9: ...viii ...
Page 16: ...Glossary Index Contents Contents 7 ...
Page 17: ...Contents 8 ...
Page 65: ...2 16 ...
Page 86: ...4 Applications ...
Page 96: ...6 Display 6 4 Scale 6 5 Offset 6 6 Input 6 6 Probe 6 7 Calibrate 6 10 Channel Menu ...
Page 98: ...Channel menu and menu map for the HP 54712A plug in Figure 6 1 Channel Menu 6 3 ...
Page 110: ...7 Thresholds 7 4 Top Base 7 6 Define time 7 8 Statistics 7 9 Define Measure Menu ...
Page 112: ...Define measure menu and menu map Figure 7 2 Define Measure Menu 7 3 ...
Page 143: ...8 22 ...
Page 163: ...9 20 ...
Page 164: ...10 Messages ...
Page 178: ...12 Function 12 3 Define Function 12 4 Display 12 7 Math Menu ...
Page 186: ...13 Measurements ...
Page 245: ...13 60 ...
Page 246: ...14 Setup Memory 14 3 Save 14 3 Recall 14 4 Default Setups 14 4 Setup Menu ...
Page 261: ...15 10 ...
Page 274: ...17 Scale 17 3 Position 17 3 Reference 17 4 Windowing 17 5 Time Base Menu ...
Page 282: ...Trigger menu and menu map Figure 18 1 Trigger Menu 18 3 ...
Page 319: ...19 18 ...
Page 322: ...20 Waveform 20 3 Pixel 20 6 Waveform Menu ...
Page 346: ...Limit test menu and menu map Figure 22 2 Limit Test Menu 22 3 ...
Page 357: ...22 14 ...
Page 360: ...Mask menu map Figure 23 1 Mask Menu 23 3 ...
Page 370: ...Figure 23 3 Figure 23 4 Mask Menu Edit Mask 23 13 ...
Page 407: ...Index 10 ...
Page 409: ......