266
43-TV-25-30 Iss.6 GLO Aug 07 UK
*Specification table for Sterilisation: The definition of Fo/Po is the sterilisation/pasteurisation time in minutes required to destroy a stated number
of organisms with a known z at temperature T. For example, "F18/250" represents the time in minutes required to destroy a stated number of
organisms at a temperature of 250°F (121.11°C) with a z = 18 degrees F. F values are used to compare the sterilizing values of different proc-
esses, however, F values cannot be compared unless the z values are the same. When temperature is not specified (for example, F = 8.6) it
is understood that the temperature is 250°F (121.11°C); the subscript O (as in the term Fo = 7.4) is used to indicate that the z = 18 degrees F
and the temperature is 250°F (121.11°C).
* The maximum rating only applies up to 104°F (40°C). For higher temperatures, the RH spec is de-rated to maintain constant moisture content
OPC Server (optional)
OPC Server (Open Process Control) - Software application for realtime interfacing between servers and
clients. OPC is a software standard that defines common interfaces for data exchange between devices
such as recorders, controllers, PCL’s and Windows
™
based applications. OPC 3.0 DA & AE compliant.
Totalisers and up to 96 pens can be transmitted via OPC server, max poll rate 1/s.
Pwd Net Sync (optional)
Password Network Synchronisation. Password can be synchronised over the network, a recorder can be
designated as a master of a password group and other recorders can be added to that group as slaves,
the master will ensure all passwords are synchronised with all recorders in its group. up to 32 recorders
can be in a Password group, this include the password Master.
Extra Pens (4 pens)
(optional)
4 extra pens to store and display totalised values, results of calculations, etc. Maximum is up to 16 extra
pens for the Minitrend QX, 12 extra pens for the eZtrend QXe recorders, and 48 extra pens for the Multi-
trend SX recorder.
Specification Environmental and Operating Conditions
Parameter
Reference
Rated
Extreme
Transport & Stor-
age
Ambient Temperature
67 °F to 77 °F
19 °C to 25 °C
32 °F to 122 °F
0 °C to 50 °C
32 °F to 122 °F
0 °C to 50 °C
14 °F to 140 °F
–10 °C to 60 °C
Relative Humidity (%RH)
50 to 65*
10 to 90*
5 to 90*
5 to 95*
Vibration
Frequency (Hz)
Acceleration (g)
0
0
0 to 70
0.1
0 to 100
0.2
0 to 100
0.5
Mechanical Shock
Acceleration (g)
Duration (ms)
0
0
1
30
5
30
20
30
Mounting Position from Vertical
Tilted Forward
Tilted Backward
Tilted to Side (+/-)
5°
5°
5°
40°
65°
65°
40°
65°
65°
Any
Any
Any
Power Requirements
Mains Voltage (Vrms)
Low Voltage AC (Vrms)
DC Voltages
Frequency (Hz)
220 to 240
24 +/- 2
24 +/- 2
49.8 to 50.2
100 to 250
20 to 30
QX/SX
20 to 25
QXe
20 to 55
QX/SX
20 to 30
QXe
47 to 63
90 to 264
20 to 30
QX/SX
20 to 25
QXe
20 to 55
QX/SX
20 to 30
QXe
47 to 63
N/A
N/A
N/A
N/A
Power Consumption
QX/QXe
: AC: <40W (max), DC: <40W (max) Typical 20W.
SX
: AC: <60W (max), DC: <60W (max) Typical 30W
Warm Up
30 minutes minimum
Seismic Qualification
Complies with IEEE 323-1974 and/or 1983 and IEEE 344-1975 and/or 1987 (option). Hard-
ware includes an IEC bracket.
Specification Options
Summary of Contents for Minitrend QX
Page 2: ...ii 43 TV 25 30 Iss 6 GLO Aug 07 UK User manual ...
Page 50: ...44 43 TV 25 30 Iss 6 GLO Aug 07 UK ...
Page 192: ...186 43 TV 25 30 Iss 6 GLO Aug 07 UK ...
Page 210: ...204 43 TV 25 30 Iss 6 GLO Aug 07 UK ...
Page 242: ...236 43 TV 25 30 Iss 6 GLO Aug 07 UK ...
Page 249: ...43 TV 25 30 Iss 6 GLO Aug 07 UK 243 Multitrend SX Recorder 14 15 32 46 50 30 31 35 36 28 55 ...
Page 250: ...244 43 TV 25 30 Iss 6 GLO Aug 07 UK 52 11 23 33 39 5 6 RS485 USB host Ethernet 1 2 3 4 ...
Page 254: ...248 43 TV 25 30 Iss 6 GLO Aug 07 UK eZtrend QXe Recorder 16 5 3 1 4 2 24 49 9 45 19 6 49 ...
Page 258: ...252 43 TV 25 30 Iss 6 GLO Aug 07 UK ...
Page 274: ...268 43 TV 25 30 Iss 6 GLO Aug 07 UK ...
Page 276: ...270 43 TV 25 30 Iss 6 GLO Aug 07 UK ...
Page 296: ...290 43 TV 25 30 Iss 6 GLO Aug 07 UK ...
Page 308: ...302 43 TV 25 30 Iss 6 GLO Aug 07 UK ...
Page 312: ...306 43 TV 25 30 Iss 6 GLO Aug 07 UK ...
Page 323: ......