50489/GTEX Chem IIC/1805
UTImeter Gtex Chem IIC
47
9.13. Storage of HERMetic devices
For a proper storage of HERMetic products
(UTImeter, Sampler, Thermometer and related
spare-parts…), we recommend:
-
Clean the devices after use,
-
Remove batteries for prolonged
storage,
-
Store batteries in a dry and cold
location,
-
Store the goods in a safe, dry and
dust free location with an ambient
temperature b5°C to +45°C.
9.14. Transportation of HERMetic
devices
For transportation of the device, always strecht out
the tape to avoid any move of the the sensor
inside its storage tube.
For transportation of the device by its handle,
always carry the unit with the button handle
directed to the body (carrier).
9.15. Recycling of HERMetic
devices
At delivery, equipment does not contain any
dangerous materials inside which can harm the
environment and people health during normal use
or disposal. However the utilization and recycling
of the equipment after the end of its life must be
implemented by an authorized organization in
accordance to local legislation.
Do not throw in rubbish but recycle wastes in
accordance to environmental / local rules.
9.16. Installation & General Care of
HERMetic Valves
Refer to Recommendation for safe use, paragraph
6.1:
...grounding is effected through the quick
connect coupler and the mating nipple of the
valve
provided that these parts are
kept
clean and free from corrosion in order to
guarantee electrical conductivity. If a
grease is used for this purpose, it must
be one which contains graphite.
For a proper installation, please refer to the
chosen sealing component manufacturer
installation guide. In any case, tightening torque of
valve fixing screws must not overtake 160 Nm
(120 lbf.ft).
Inspect valves in regards to damage / marks /
pollution preventing a proper connection and
gastightness when connected with the HERMetic
devices.
Ensure no damage impact the PTFE sealing of
the ball valve.
Where appropriate, complete a leak test with leak
detector spray to confirm the valve tightness.
Summary of Contents for HERMetic UTImeter Gtex Chem IIC
Page 5: ...50489 GTEX Chem IIC 1805 5 UTImeter Gtex Chem IIC Figure 2 2 ...
Page 15: ...50489 GTEX Chem IIC 1805 UTImeter Gtex Chem IIC 15 Figure 7 2 ...
Page 63: ......
Page 64: ......
Page 65: ......
Page 66: ......
Page 67: ......
Page 68: ......
Page 69: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ......
Page 80: ......