50489/GTEX Chem IIC/1805
UTImeter Gtex Chem IIC
46
9.11. Temperature verification
The temperature calibration curve is stored in the
sensor memory and cannot be modified. The
calibration is set once at the factory and do not
require subsequent adjustment.
Nevertheless it is recommended to check the
temperature accuracy once a year. A one point
check is enough to qualify the sensor.
9.11.1. Equipment required
-
A Dewar flask or any vacuum flask,
approximately 8 cm in diameter and 36 cm
deep.
-
Ice, preferably made from distilled water.
-
Water, preferably distilled and precooled.
9.11.2. Preparing the Ice Point bath
-
Shave or crush the ice into small pieces,
avoiding direct contact with the hands or any
unclean object. The pieces shall be no more
then 5 mm.
-
Fill the Dewar flask with the crushed ice and
add sufficient water to form a slush, just filling
the voids between ice particles but not enough
to float the ice.
-
Insert the sensor, packing the ice gently about
it.
-
Let it stand for half an hour to permit the
sensor temperature, the ice particles and the
water to equilibrate.
-
As the ice melts it will be necessary to drain
off some water and add more crushed ice.
Gently stir the ice with the sensor periodically
to assist equilibration.
IMPORTANT NOTE
: Attention to detail during the
preparation of the Ice Point bath is critical to
the accuracy and quality of the offset
verification.
5.5.1
Checking the UTImeter
-
After 30 minutes have elapsed, gently stir the
bath with the sensor again to ensure complete
equilibration of temperature.
-
Switch on the UTImeter.
-
Observe the reading. It should be
±
0.10 °C
(
±
0.20 °F) The temperature must be stable,
i.e. within
±
0.04 °C (
±
0.07 °F).
-
If it is not OK, refer to section 11 "Trouble
shooting”
9.12. Ullage/Interface verification
The sensitivity of the instrument in ullage /
interface cannot be adjusted. Both ullage and
interface levels are set at the factory.
Checking ullage and interface level detection
The test liquid should be the one to be gauged. Fill
in a container with appropriate liquid.
Switch on the unit. The buzzer shall beep every 2
sec.
If the liquid is conductive (alcohol, water, ...)
- Check the
ullage
by immersing the ultrasonic
gap sensor but not the electrodes (position A);
The buzzer shall beep continuously.
- Check the
interface
by immersing the
interface electrodes (position B). The buzzer
shall beep intermittently.
If the liquid is non conductive (gasoline, oil, ...)
- Check the
ullage
by immersing the sensor
(position B); The buzzer shall beep
continuously.
- Check the interface by immersing the sensor
(position B) in water. The buzzer shall beep
intermittently.
position A
position B
Figure 10-12
Summary of Contents for HERMetic UTImeter Gtex Chem IIC
Page 5: ...50489 GTEX Chem IIC 1805 5 UTImeter Gtex Chem IIC Figure 2 2 ...
Page 15: ...50489 GTEX Chem IIC 1805 UTImeter Gtex Chem IIC 15 Figure 7 2 ...
Page 63: ......
Page 64: ......
Page 65: ......
Page 66: ......
Page 67: ......
Page 68: ......
Page 69: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ......
Page 80: ......