9.4. Sensing probe replacement
THE REPLACEMENT OF THE SENSING
PROBE DOES NOT REQUIRE TO RE-
CALIBRATE THE TEMPERATURE NOR THE
ULLAGE / INTERFACE.
9.4.1. Disconnecting the old sensing probe
−
Unscrew the securing screw with the 1.5 mm
Hex Allen key.
−
Pull carefully the adaptor out of the sensing
probe tube by turning it slightly left and right.
Make sure that the O-ring is not damaged
when it passes the hole of the sensing probe
tube.
−
Disconnect the plug by pulling it gently out of
the tube.
9.4.2. Connecting the new sensing probe
−
Refer to Figure 10-8.
−
Insert the Insertion tool gently in the free
external holes of the tape plug.
−
With one hand keep the sensing probe and the
tape adaptor as shown on Figure 10-8.
−
With the other hand drive the plug into the new
sensor tube with the Insertion tool to connect it
to the sensing probe socket. Note this is a one
way only plug. The wires shall be on the
opposite side of the electronic circuit print as
shown on Figure 10-8.
−
Pull out gently the Insertion tool from the plug.
If needed, keep the plug in place with another
non sharp tool, for instance the 4 mm Allen
key. Check that the plug is fully inserted.
−
Switch on the unit and wait a few seconds. If all
is OK, the temperature is displayed and the
buzzer beeps every 2 seconds. If there is any
problem, refer to the section 11 "Trouble
−
Put some light grease on the O-ring.
−
Push gently the adaptor into the sensing probe
tube. Mind not to damage the O-ring when it
passes the screw hole.
−
Screw the securing screw back with the 1.5
mm Hex Allen key.
Figure 10-8
9.5. Tape wipers replacement and
removing of tape cover
The 2 tape wipers can be easily replaced:
−
Check that the tape cleaner is on "DOWN"
position.
−
Pull the tape cover out of the frame. Use pliers
or a rod to help the clips to get out of the frame
(as shown in figure 9.10)
−
The tape wipers are inserted in holders
grooves. Remove the old ones and insert the
new ones.
−
Push the tape cover back into the frame.
−
Check that the tape cleaner is working
properly.
Note:
we recommend to change always both
wipers.
wipers
Figure 10-9
Summary of Contents for HERMetic UTImeter Gtex Chem IIC
Page 5: ...50489 GTEX Chem IIC 1805 5 UTImeter Gtex Chem IIC Figure 2 2 ...
Page 15: ...50489 GTEX Chem IIC 1805 UTImeter Gtex Chem IIC 15 Figure 7 2 ...
Page 63: ......
Page 64: ......
Page 65: ......
Page 66: ......
Page 67: ......
Page 68: ......
Page 69: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ......
Page 80: ......