50489/GTEX Chem IIC/1805
UTImeter Gtex Chem IIC
11
5. Recommendation for safe use
According to TSB_7030_E, Issue 1 of November 18, 2014
1. This Operation and Service Manual is a guide in order to help the user to operate the device safely and
correctly.
2. Nevertheless the maker disclaims all responsibility and liability for damage resulting from the use of the
equipment regardless of the cause of the damage.
3.
Before using the device, ensure the device fits with the gauging conditions (tank pressure,
product, temperature…). Refer to device specifications.
4.
This device is certified to penetrate into a “Zone 0” (explosible area) when connected to a valve.
The opening of the valve may generate risk of flammable gas release or flame entrance. Ensure
safety conditions are met before use.
5.
Attention is drawn to the possible hazard due to electrostatic charges which may be present in the
tank.
This may happen in particular with static accumulator liquids, i.e. liquids which have low conductivity
of 50 picoSiemens/metre (pS/m) or less.
6.
It is very important that the instrument is grounded to the tank before the probe is introduced into
the tank and remains grounded until after complete withdrawal from the tank.
6.1. If the instrument is installed with the quick connect coupler, grounding is effected through the quick
connect coupler and the mating nipple of the valve provided that these parts are kept clean and free
from corrosion in order to guarantee electrical conductivity. If a grease is used for this purpose, it
must be one which contains graphite.
6.2. If the instrument is not connected to the mating deck valve, the instrument has to be also earthed by
means of the grounding cable and clamp.
7.
It is anticipated that the user will have specific operating methods laid down to ensure safety
when using this type of apparatus. In this case the user's instructions shall be strictly observed.
8.
In the absence of such instructions the following should be noted:
8.1. If a metal sounding pipe is fitted beneath the deck valve or tank is inerted, then ullaging, etc. is
permissible at any time with no restriction.
8.2. If there is no sounding tube or tank is not inerted, the following precautions shall be taken:
8.2.1. If the cargo is not a static accumulator liquid, i.e. its conductivity is more than 50 pS/m, then
ullaging is permitted provided that the instrument is properly grounded and earthed before
the probe is inserted into the tank and remains earthed until the probe has been removed
from the tank.
8.2.2. If the cargo is a static accumulator liquid, i.e. its conductivity is less than 50 pS/m, then
ullaging is permitted provided that:
8.2.2.1. The instrument is properly grounded and earthed before the probe is inserted into
the tank and remains earthed until the probe has been removed from the tank.
8.2.2.2. The apparatus is not introduced into a tank until at least 30 minutes have elapsed
after completion of any loading operation or stopping the injection of inert gas.
Summary of Contents for HERMetic UTImeter Gtex Chem IIC
Page 5: ...50489 GTEX Chem IIC 1805 5 UTImeter Gtex Chem IIC Figure 2 2 ...
Page 15: ...50489 GTEX Chem IIC 1805 UTImeter Gtex Chem IIC 15 Figure 7 2 ...
Page 63: ......
Page 64: ......
Page 65: ......
Page 66: ......
Page 67: ......
Page 68: ......
Page 69: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ......
Page 80: ......