9.3. Tape replacement
THE REPLACEMENT OF THE TAPE DOES NOT REQUIRE TO RE-CALIBRATE THE TEMPERATURE.
Follow the different sequences as described below. The Figure 13-1 : general assembly, list of the main
spare parts can also help.
9.3.1. Disconnecting the tape from the
sensor
Follow the instructions of section 10.4 "Sensing
9.3.2. Disconnecting the tape from the
electronic box
−
Unscrew with the 2.5 Allen key the 2 screws
(A) of the battery holder and pull it out as
shown on Figure 10-4.
−
Unscrew with the 2.5 Allen key the 4 screws
(B) of the display unit and pull it gently out as
shown on Figure 10-4.
−
Disconnect the connecting plug (C) as shown
on Figure 10-5 and remove the display unit.
−
Unscrew with the 2.5 Allen key the tape holder
(G) by removing the 2 screws (F) and the
grounding cable (D) as shown on Figure 10-5.
Do not loose the 2 remaining screws that
secure the reel axle.
(A)
(B)
Figure 10-4
(D) grounding cable of tape
(E) black oversleeve of tape
(G) tape holder
(H) tape
(C) connecting plug
to display unit
(I) securing screw
(F) securing screw
Figure 10-5
Summary of Contents for HERMetic UTImeter Gtex Chem IIC
Page 5: ...50489 GTEX Chem IIC 1805 5 UTImeter Gtex Chem IIC Figure 2 2 ...
Page 15: ...50489 GTEX Chem IIC 1805 UTImeter Gtex Chem IIC 15 Figure 7 2 ...
Page 63: ......
Page 64: ......
Page 65: ......
Page 66: ......
Page 67: ......
Page 68: ......
Page 69: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ......
Page 80: ......