Guidelines and manufacturer's statement - electromagnetic compatibility
41
EN
b
The field strength should be less than 3 V/m for the frequency range of 150 kHz
to 80 MHz.
11.6
Recommended safety distances between portable and mobile
HF telecommunications devices and the Aquilex Fluid Control
System
The Aquilex Fluid Control System is intended for use in an electromagnetic envi-
ronment where HF interferences are controlled. The user/operator of the Aquilex
Fluid Control System can contribute to lowering electromagnetic emissions by
complying with the minimum distance between portable and mobile HF tele-
communications devices (transmitters) and the Aquilex Fluid Control System -
depending on the output power of the communication device listed below.
The safety distance d in meters [m] for transmitters with a max. rated output not
listed in the table above can be calculated by applying the corresponding formula
in the respective column. P is the max. rated output of the transmitter in watts
[W] according to the information provided by the manufacturer of the transmit-
ter.
Note 1: The higher frequency range applies to 80 and 800 MHz.
Note 2: These guidelines are probably not realizable in all cases. The distribution
and spread of electromagnetic quantities differs depending on the absorption
and reflection of buildings, objects, and people.
Rated output
of the trans-
mitter [W]
Safety distance based on the transmitting frequency [m]
150 kHz to 80 MHz
d = 1.2
P
80 MHz to 800 MHz
d = 1.2
P
800 MHz to 2.5 GHz
d = 2.3P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
Summary of Contents for AQL-100
Page 6: ......
Page 50: ...46 Appendix EN 15 Appendix 15 1 Test Log Date Tests Performed Results Comment Signature ...
Page 52: ......
Page 54: ......
Page 102: ......
Page 104: ......
Page 152: ......
Page 154: ......
Page 202: ......
Page 204: ......
Page 252: ......
Page 254: ......
Page 302: ......