Tubing Overview
Three different tube sets are necessary to operate the system. The following ta-
ble lists each type of tube set and its application.
Table 5-1
5.8
Connecting the Outflow Tubing
CAUTION!
When using the Aquilex System with MyoSure® or other morcellation systems,
the combination of low set pressures and high vacuum pressures may result in a
significant loss of intrauterine distension pressure which has the potential to af-
fect the visibility of the surgical field. Conversely, when employing high disten-
sion pressures, the deactivation of the MyoSure® or other morcellator system
can lead to pressure spikes that can exceed 150 mmHg. These situations may oc-
cur for a short time as the system automatically adjusts the flow rate to return
to the set intrauterine pressure.
Using Low Vacuum configuration of Figure 5-4, connect outflow tubing (Y-tube)
to patient port
(4)
of second canister. Yellow flexible connector attaches to
drape
(6)
. Yellow luer fitting connects to stopcock
(7)
of Removable Outflow
Part Number Description
AQL-110
Aquilex Fluid Control System Inflow Tube Set
AQL-111
Aquilex Fluid Control System Outflow Tube Set
AQL-112
Complete tube set (Inflow and Outflow) disposable, sterile
AQL-114
Aquilex Fluid Control System High and Low Vacuum Tube
Set: re-usable, non-sterile
Figure 5-7 Outflow Tubing
(1)
To low vacuum port (white)
(2)
Canisters
(3)
Tandem tube
(4)
Patient port
(5)
Outflow tube set
(6)
Drape
(7)
Removable Outflow Channel (Myo-
Sure®) or hysteroscope outflow
sheath stopcock
(2)
(6)
(3)
(4)
(5)
(1)
(7)
(5)
(3)
(1)
Summary of Contents for AQL-100
Page 6: ......
Page 50: ...46 Appendix EN 15 Appendix 15 1 Test Log Date Tests Performed Results Comment Signature ...
Page 52: ......
Page 54: ......
Page 102: ......
Page 104: ......
Page 152: ......
Page 154: ......
Page 202: ......
Page 204: ......
Page 252: ......
Page 254: ......
Page 302: ......