48
Chapter 4: Commissioning
4. If the calibration was not successful, “Error”
will appear.
In the event of an error, check the entire sys-
tem. Resistance in the overall system may be
too high. Then rerun the calibration.
ST
Tuesday l 22.03.2017 l 20:36
40
30
20
10
0
Calibrate Tube
100 l/min
Error
Press the MFK to continue
If you do not perform a tube calibration, the calibration data from the last tube calibra-
tion are used. Therapy can be started as usual, but the pressure measurement may be
incorrect. We recommend always performing tube calibration.
When commissioning the device for the first time without calibrating it, the default cali-
bration data
1
stored in the device are used.
1
Operation without humidifier:
Leakage tube circuit:
∅
22 mm; 1.80 m
Height: ca. 44 m NHN, air pressure: approx. 1008 hPa
Connected accessories: no
Operation with humidifier:
Leakage tube circuit:
∅
22 mm; 1.80 m
Height: ca. 44 m NHN, air pressure: approx. 1008 hPa
Connected accessories: Humidifier AquaTREND III-NG
Summary of Contents for 5CPD10
Page 2: ......
Page 10: ...10 List of figures ...
Page 11: ...Chapter 1 Introduction 11 Chapter 1 Introduction ...
Page 21: ...Chapter 2 Safety Information 21 Chapter 2 Safety Information ...
Page 28: ...28 Chapter 2 Safety Information ...
Page 29: ...Chapter 3 Description of Device 29 Chapter 3 Description of Device ...
Page 36: ...36 Chapter 3 Description of Device ...
Page 37: ...Chapter 4 Commissioning 37 Chapter 4 Commissioning ...
Page 56: ...56 Chapter 4 Commissioning ...
Page 57: ...Chapter 5 Therapy Modes 57 Chapter 5 Therapy Modes ...
Page 65: ...Chapter 6 Operating the Device 65 Chapter 6 Operating the Device ...
Page 91: ...Chapter 7 Notifications Messages 91 Chapter 7 Notifications Messages ...
Page 95: ...Chapter 8 Cleaning and Disinfection 95 Chapter 8 Cleaning and Disinfection ...
Page 104: ...104 Chapter 8 Cleaning and Disinfection ...