Chapter 8: Cleaning and Disinfection
Using the device for more than one patient
If the device is intended for use by more than one patient (e.g., operation in clinics), a
suitable bacterial filter (e.g. Medisize Barr-vent S) should be used continuously to pro-
tect the device from contamination by human pathogens. The bacterial filter must be
changed daily.
Risk of infection due to germs!
•
If the device has not been continuously protected by a bacterial
filter, it may be contaminated with pathogenic germs. If the
device is used by another patient, the patient could become
infected with these germs.
Ö
Protect the device continuously using a bacterial filter.
Ö
Thoroughly clean and disinfect the device between use by
different patients to ensure it is free of pathogenic germs.
Ö
If in doubt, always assume that the device is contami-
nated, and has to be hygienically reconditioned accord-
ing to the hygienic concept.
Ö
If
contamination with multi-resistant pathogens (e.g.
MRSA)
is suspected, the device must be packaged with
the appropriate labelling and disinfected accordingly.
•
Accessories and consumable materials (e.g. tubes, mask etc.)
can be contaminated with human pathogens. If they are
used by another patient, the patient could become infected
with these germs.
Ö
Accessories and consumable materials labelled as “single
use”
may only be used once for one patient.
Ö
Accessories and consumable materials labelled as “sin-
gle patient use” may only be used for one patient and
must be cleaned in accordance with the manufacturer’s
instructions.
Summary of Contents for 5CPD10
Page 2: ......
Page 10: ...10 List of figures ...
Page 11: ...Chapter 1 Introduction 11 Chapter 1 Introduction ...
Page 21: ...Chapter 2 Safety Information 21 Chapter 2 Safety Information ...
Page 28: ...28 Chapter 2 Safety Information ...
Page 29: ...Chapter 3 Description of Device 29 Chapter 3 Description of Device ...
Page 36: ...36 Chapter 3 Description of Device ...
Page 37: ...Chapter 4 Commissioning 37 Chapter 4 Commissioning ...
Page 56: ...56 Chapter 4 Commissioning ...
Page 57: ...Chapter 5 Therapy Modes 57 Chapter 5 Therapy Modes ...
Page 65: ...Chapter 6 Operating the Device 65 Chapter 6 Operating the Device ...
Page 91: ...Chapter 7 Notifications Messages 91 Chapter 7 Notifications Messages ...
Page 95: ...Chapter 8 Cleaning and Disinfection 95 Chapter 8 Cleaning and Disinfection ...
Page 104: ...104 Chapter 8 Cleaning and Disinfection ...