128
Chapter 9: Annex
Guidance and manufacturer’s declaration - electromagnetic immunity
Immunity tests
IEC 60601 test level
Compliance level
Electromagnetic envi-
ronment - guidance
Voltage dips, tem-
porary interruptions
and voltage fluctua-
tions in the voltage
supply lines as per
IEC 61000-4-11
> 95 % dip of
U
T
for
1/2 period
> 95 % dip of
U
T
for 1
period
> 30 % dip of
U
T
for
25 (50 Hz) periods/
30 (60 Hz) periods/
> 95 % dip of
U
T
for
250 (50 Hz) periods /
300 (60 Hz) periods /
> 95 % dip of
U
T
for
1/2 period
> 60 % dip of
U
T
for
5 periods
> 30 % dip of
U
T
for
25 periods
> 95 % dip of
U
T
for
5s or 6s
The quality of the sup-
ply voltage should cor-
respond to one of the
typical business or hos-
pital environments. If
the user of the
TREND
III
respiratory therapy
device requires contin-
ued functioning even in
the event of an interrup-
tion to the power supply,
we recommended supply-
ing the
TREND III
respira-
tory therapy device using
an uninterruptable power
supply (UPS) or a battery.
Summary of Contents for 5CPD10
Page 2: ......
Page 10: ...10 List of figures ...
Page 11: ...Chapter 1 Introduction 11 Chapter 1 Introduction ...
Page 21: ...Chapter 2 Safety Information 21 Chapter 2 Safety Information ...
Page 28: ...28 Chapter 2 Safety Information ...
Page 29: ...Chapter 3 Description of Device 29 Chapter 3 Description of Device ...
Page 36: ...36 Chapter 3 Description of Device ...
Page 37: ...Chapter 4 Commissioning 37 Chapter 4 Commissioning ...
Page 56: ...56 Chapter 4 Commissioning ...
Page 57: ...Chapter 5 Therapy Modes 57 Chapter 5 Therapy Modes ...
Page 65: ...Chapter 6 Operating the Device 65 Chapter 6 Operating the Device ...
Page 91: ...Chapter 7 Notifications Messages 91 Chapter 7 Notifications Messages ...
Page 95: ...Chapter 8 Cleaning and Disinfection 95 Chapter 8 Cleaning and Disinfection ...
Page 104: ...104 Chapter 8 Cleaning and Disinfection ...