(2)Constant Pressure Mode
This section explains the setting of liquid delivery system at
constant pressure. An example is given where initial flow rate is
1.0 mL/min, at a pressure of 15.0 MPa.
Monitor screen
MANUAL
MANUAL:
FLOW=1, PRESS=2 [MN]
2
Select the constant pressure mode.
CONSTANT PRESS: INITIAL FLOW PRESS [MN]
1.000 15.0
Input the flow rate, and the pressure values.
CONSTANT FLOW:
FLOW PRESS [MN]
0.000 0.0
The monitor screen will apper.
Monitor screen
Before stating the pump, check that the pressure level is 0 Mpa.
NOTICE:
The liquid delivery will not start unless the starting pressure
is 0 MPa
NOTICE:
The set conditions for manual operation will be erased when
power is turned off. Conditions must be set each time power is turned
on.