1.2 Low-Pressure Gradient System
In the low-pressure gradient system, two or more solvents are
mixed in its low-pressure section from which a liquid feed is
performed by a single pump. The composition of an eluent is
made to vary with time in chromatographic separation.
(1) Each of plural eluents is fed from the degasser to the low-
pressure unit.
(2) In the low-pressure unit, the eluents are mixed at a pre-
specified ratio.
(3) Through the mixer, the mixture eluent is delivered.
(4) A sample is injected by the autosampler.
(5) The injected sample is separated through the column which
is maintained at a constant temperature by the column
oven.
(6) Each component separated from the sample is detected as
a signal by the detector.
1.2
Model L-2100/2130 pump
Model L-2300 column oven
Model L-2200 autosampler
Model L-2400 UV detector
Degasser
Washing solution
Mixer
Low-pressure
gradient unit
Eluent
System power supply unit
Drain
Column