Table 3-1 Name and Function of Each Terminal at Rear Panel
Classi-
fication
Name
Function
Contents of
Signal
Contents of
Processing
Remarks
TIMER
(1 to 4)
Contact
turns
on/off
according
to coded
setting in
EVENT of
the
program.
•
Timer
ON: CLOSE
OFF: OPEN
•
Pulse
OPEN
The ON/OFF
signal can
output to
each TIMER by
specifying an
ON/OFF EVENT
code at a
desired time
of a time
program.
1-second
CLOSE signal
can output at
each TIMER by
specifying
pulse EVENT
code at a
desired time
of a time
program.
•
TIMER No. and Event code
•
Pump can be stopped by specifying
00 for EVENT, and buzzer can be
rung by specifying 92.
1) Becomes OFF when the pulse is
output in the ON status.
2) OFF has priority over ON.
3) Contact output specification of
each terminal
Voltage ... 30 V DC max.
Current ... 100 mA max.
Output
PRESS
OUT
Pressure
value is
output to
the
exterior.
0 to 1 V DC
•
1 V = Approx. 49 MPa
(Approx. 7106 psi)
PUMP
ON IN
Pump ON
from
external
system.
Input
PUMP
OFF IN
Pump OFF
from
external
system.
OPEN
∗
50 ms
or longer
Maximum current connectable: 10 mA
∗
Edge jugement
Signal
TIMER
No.
OFF
ON
Pulse
1
10
11
12
2
20
21
22
3
30
31
32
4
40
41
42
CLOSE
1 s
CLOSE
OPEN
3.10