2 - 4
2.3 Items to be Prepared by the User
(1) Power voltage : 100 to 240 V AC
Fluctuation .......Within
±
10% with respect
to rated voltage
(2) Frequency
: 50 or 60 Hz
Fluctuation .......Within
±
0.5 Hz with respect
to rated frequency
(3) Power
consumption : 500 VA or more in total system
Provide power supply sufficient for
combinational use with other devices.
(4) Grounding
: Grounding connection of Electric
Installation Technical Standard Class 3 or
higher should be provided.
WARNING
Ground Properly to Prevent Electric Shock Hazard!
•
Be sure to use the power cable supplied with the
instrument. Use of a different power cable could result in
an electric shock hazard.
•
This instrument is of a "plug-connected type" specified in
European Norm EN61010-1, so connect the power cable to
a grounded three-wire outlet.
•
If a grounded three-wire outlet is not available, the power
cable may be connected to a two-wire outlet through an
adapter plug. In this case, be sure to provide proper
grounding connection as required.
2.3 Items to be Prepared by the User
2.3.1 Power Supply