4 - 5
4.1.3
Manual Operation
This is used for supplying liquid under constant conditions
without using a program.
(1)Constant Flow Rate Mode
This section explains the setting of liquid delivery system at
constant flow rate. An example is given where flow rate at 1.0
mL/min, upper pressure limit at 15.0 MPa and lower pressure
limit at 0 MPa.
Monitor screen
MANUAL
MANUAL:
FLOW=1, PRESS=2 [MN]
1
Select the constant flow rate mode.
CONSTANT FLOW:
FLOW MAX MIN [MN]
1.000 15.0 0.0
Input the flow rate, and the upper and lower limit pressure
values.
CONSTANT FLOW:
FLOW PRESS [MN]
0.000 0.0
The monitor screen will apper.
Monitor screen
NOTICE:
The set conditions for manual operation will be erased
when power is turned off. Conditions must be set each time power
is turned on.