4 - 36
4.4.2
Pressure Monitor
The number of double-speed section errors in each measurement can be indicated.
A double-speed section error indicates that the pressure did not return to the reference level
(previous pressure) even after double-speed control (an increase of pump motor speed) was
effected to compensate for a pressure drop. This may be caused by formation of air
bubbles in the flow path.
Monitor screen
GLP`
GLP:
LOGBOOK=1, PRESSURE M.=2,
[--]
2
KEY LOCK=
Select PRESSURE M.
PRESSURE MONITOR
[--]
9999
TIMES
Display the number of double-speed section errors.
ESC
Monitor screen
Resetting occurs on the following occasions.
•
On input of [PUMP ON] or [PUMP OFF]
•
On activation of contact
•
On start of gradient elution
•
On reception of start signal from integrator in D-line configuration
•
On external start of pump
•
On input of [PURGE ON] or [PURGE OFF]