4 - 28
FLUSH MODE: PARAM.=1, PROG.(NO.9)=2
[--]
1
Enter “1” for column flushing under constant conditions, or “2” for column flushing
using a gradient.
[PARAM]
SET DELAY TIME:
TIME (min)
[--]
<1.0-600.0>
600.0
Input the time period from the end of analysis to the
start of column flushing.
COLUMN FLUSH:
FLOW
[--]
1.000
Input column flushing conditions (flow rate).
FLUSHING TIME:
TIME (min)
[--]
<1.0-600.0>
30.0
Column flushing time can be input within 1.0 to 600.0
min.
Monitor screen
[NO]
Monitor screen
When the time for column flushing arrives, the flushing is
executed according to No.9 program.
NOTICE:
The program for column flushing is fixed at No. 9.
Upon selecting column flushing, program No. 9 is
automatically set for flushing.
When column flushing is set, program No. 9 is not
usable as a gradient program. But program No. 9 is
usable with a link program.
The pump is stopped automatically when program
No. 9 ends.